VITAL SIGNS
Received Sep 11, 2025 · Event occurred Aug 12, 2025
Report 1314417-2025-00068 · MDR key 23032572
Device
Generic name
Vital Signs¿ Adult Anesthesia Breathing Circuit
Manufacturer
Vyaire Medical Inc.Model number
AMY520X4FCatalog number
AMY520X4F
Lot number
UNKNOWN
Product problems
- Partial Blockage
- No Device Output
- Failure to Deliver
- Blocked Connection
- No Flow
- Partial Blockage
- No Device Output
- Failure to Deliver
- Blocked Connection
- No Flow
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
THE PRODUCT INVOLVED IN THE REPORT HAS NOT BEEN RETURNED. A REVIEW OF THE DEVICE HISTORY RECORD IS IN-PROGRESS. ALL INFORMATION REASONABLY KNOWN AS OF 11 SEP 2025 HAS BEEN INCLUDED IN THE HEALTH AUTHORITY REPORT. SHOULD ADDITIONAL INFORMATION BE OBTAINED A FOLLOW-UP HEALTH AUTHORITY REPORT WILL BE PROVIDED. THE INFORMATION PROVIDED BY AIRLIFE. REPRESENTS ALL OF THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT/REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO AIRLIFE. AIRLIFE HAS NO INDEPENDENT KNOWLEDGE OF THE EVENT REPORTED BUT IS RELAYING THE INFORMATION THAT WAS PROVIDED BY THE USER FACILITY WHERE THE INCIDENT OCCURRED. THIS PRODUCT INCIDENT IS DOCUMENTED IN THE AIRLIFE COMPLAINT DATABASE AND IS IDENTIFIED AS COMPLAINT (B)(4). THIS INFORMATION IS SUBMITTED PURSUANT TO 21 CFR 803, IN COMPLIANCE WITH THE MEDICAL DEVICE REPORTING REQUIREMENT AND SHOULD NOT BE CONSIDERED TO BE AN ADMISSION THAT A AIRLIFE PRODUCT IS DEFECTIVE OR CAUSES SERIOUS INJURY. THE COMPLAINT OF "OCCLUDED" REGARDING PART AMY520X4F WAS CONFIRMED. THE ROOT CAUSE WAS DETERMINED AS MATERIAL. A RISK ASSESSMENT WAS PERFORMED, AND THE ULTIMATE RISK WAS DETERMINED TO BE LOW WHICH DOES NOT REQUIRE ESCALATION TO THE CAPA REVIEW BOARD. THERE HAVE BEEN 1 OTHER COMPLAINTS REGARDING THE SAME PART AND A SIMILAR ISSUE WITHIN THE 24 MONTHS PRECEDING THIS REPORTED EVENT.
Description of Event or Problem
IT WAS REPORTED THAT DURING PROCEDURE THE HME PORT WAS OCCLUDED. WHILE NO HARM OR INJURY WAS REPORTED TO THE PATIENT IF THIS WERE TO REOCCUR THE PATIENT COULD BECOME HYPOXIC.
Additional Manufacturer Narrative
THE PRODUCT INVOLVED IN THE REPORT HAS NOT BEEN RETURNED. A REVIEW OF THE DEVICE HISTORY RECORD IS IN-PROGRESS. ALL INFORMATION REASONABLY KNOWN AS OF 11 SEP 2025 HAS BEEN INCLUDED IN THE HEALTH AUTHORITY REPORT. SHOULD ADDITIONAL INFORMATION BE OBTAINED A FOLLOW-UP HEALTH AUTHORITY REPORT WILL BE PROVIDED. THE INFORMATION PROVIDED BY AIRLIFE. REPRESENTS ALL OF THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT/REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO AIRLIFE. AIRLIFE HAS NO INDEPENDENT KNOWLEDGE OF THE EVENT REPORTED BUT IS RELAYING THE INFORMATION THAT WAS PROVIDED BY THE USER FACILITY WHERE THE INCIDENT OCCURRED. THIS PRODUCT INCIDENT IS DOCUMENTED IN THE AIRLIFE COMPLAINT DATABASE AND IS IDENTIFIED AS COMPLAINT (B)(4). THIS INFORMATION IS SUBMITTED PURSUANT TO 21CFR803, IN COMPLIANCE WITH THE MEDICAL DEVICE REPORTING REQUIREMENT AND SHOULD NOT BE CONSIDERED TO BE AN ADMISSION THAT A AIRLIFE PRODUCT IS DEFECTIVE OR CAUSES SERIOUS INJURY.
Description of Event or Problem
IT WAS REPORTED THAT DURING PROCEDURE THE HME PORT WAS OCCLUDED. WHILE NO HARM OR INJURY WAS REPORTED TO THE PATIENT IF THIS WERE TO REOCCUR THE PATIENT COULD BECOME HYPOXIC.