CENTARGO CT INJECTOR
Received Sep 11, 2025 · Event occurred Aug 20, 2025
Report 2520313-2025-00024 · MDR key 23026326
Device
Generic name
Ct Injection System,
Manufacturer
Imaxeon Pty LtdModel number
88628624Catalog number
CENT-SYS-BAT
Product problems
- Adverse Event Without Identified Device or Use Problem
- Insufficient Device Problem Information
- Adverse Event Without Identified Device or Use Problem
- Insufficient Device Problem Information
Patient
66 YR · Female
- Air Embolism
- Air Embolism
Narrative
Additional Manufacturer Narrative
BAYER SERVICE PERFORMED A SYSTEM SERVICE CHECK OUT OF THE MEDRAD® CENTARGO CT INJECTION SYSTEM (SERIAL NUMBER (B)(6)) ON (B)(6) 2025 WHICH CONFIRMED THAT THE EQUIPMENT WAS OPERATING TO SPECIFICATION. THE CUSTOMER DISCARDED THE DISPOSABLES THAT WERE IN USE AT THE TIME OF THE INCIDENT; HOWEVER, THEY WERE ABLE TO PROVIDE THE LOT NUMBER FOR THE DAY SET AND PATIENT LINE. BAYER PRODUCT ANALYSIS FUNCTIONALLY TESTED A RETAINED SAMPLE FROM CENT-DS, LOT NUMBER 1498374001, AND A RETAINED CENT-PL, LOT NUMBER 250303. FUNCTIONAL TESTING CONCLUDED THAT THE RETAINED DISPOSABLE PERFORMED TO SPECIFICATION WITH NO PROBLEMS OBSERVED. THE MEDRAD® CENTARGO CT INJECTION SYSTEM OPERATION MANUAL CAUTIONS THE USER AS FOLLOWS: THE SYSTEM MUST BE ARMED PRIOR TO PERFORMING ANY INJECTION IN A PROTOCOL. PRESS ARM TO ARM THE SYSTEM. FOR THE FIRST INJECTION OF AN EXAM, A MESSAGE DISPLAYS ASKING FOR CONFIRMATION THE PATIENT LINE HAS BEEN CHECKED FOR AIR. PRESS YES TO CONFIRM ALL AIR HAS BEEN EXPELLED AND NO AIR IS VISIBLE IN THE PATIENT LINE PRESS NO IF THE PATIENT LINE HAS NOT BEEN CHECKED FOR AIR. THE SYSTEM WILL NOT ARM. WARNING: AIR EMBOLISM HAZARD - SERIOUS PATIENT INJURY OR DEATH MAY RESULT. DO NOT MODIFY OR ATTEMPT TO CIRCUMVENT THE OPERATION OF THE AIR DETECTORS. DO NOT CONNECT THE PATIENT LINE TO THE PATIENT UNTIL ALL TRAPPED AIR HAS BEEN CLEARED. THE CUSTOMER CONTINUES TO USE THE MEDRAD® CENTARGO CT INJECTION SYSTEM WITH NO FURTHER ISSUES REPORTED AND HAS DECLINED THE OFFER OF ADDITIONAL CLINICAL APPLICATIONS TRAINING. THIS INFORMATION DOES NOT CONSTITUTE AN ADMISSION THAT THE DEVICE, THE COMPANY, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO A REPORTABLE EVENT.
Description of Event or Problem
BAYER MEDICAL CARE WAS NOTIFIED OF AN ALLEGED AIR INJECTION THAT HAD OCCURRED DURING A CT SCAN WHILE A PATIENT WAS CONNECTED TO A MEDRAD® CENTARGO CT INJECTION SYSTEM (SERIAL NUMBER (B)(6)). FOLLOWING THE INJECTION, THE CUSTOMER REPORTED THAT THE QUANTITY OF AIR THAT WAS VISUALIZED ON THE DISPLAYED IMAGES COULD NOT BE QUANTIFIED. THE PATIENT WAS PLACED INTO A 2-HOUR HYPERBARIC CHAMBER PROTOCOL AFTER WHICH THEY WERE PLACED ON A HIGH CONCENTRATION OXYGEN MASK AND ADMITTED TO THE HOSPITAL FOR FURTHER CARE AND TREATMENT. THE CURRENT STATUS OF THE PATIENT IS UNKNOWN.
Additional Manufacturer Narrative
A SYSTEM SERVICE CHECK OF THE MEDRAD® CENTARGO CT INJECTION SYSTEM (SN (B)(6)) WAS COMPLETED ON (B)(6) 2025 WHICH CONFIRMED THAT THE INJECTOR WAS OPERATING WITHIN SPECIFICATIONS. THE DISPOSABLES THAT WERE IN USE DURING THE EVENT WERE DISCARDED BY THE CUSTOMER. HOWEVER, AS THE CUSTOMER DID PROVIDE THE LOT NUMBER FOR THE DAY SET (CENT-DS, LOT NUMBER 1498374001) AND PATIENT LINE (CENT-PL, LOT NUMBER 250303) THAT WERE IN USE, TESTING OF A RETAINED SAMPLE OF EACH IS PENDING. ONCE THE INVESTIGATION IS COMPLETED, A FOLLOW-UP REPORT WILL BE SUBMITTED. THE CUSTOMER CONTINUES TO USE THE MEDRAD® CENTARGO CT INJECTION SYSTEM WITH NO FURTHER ISSUES REPORTED AND HAS DECLINED THE OFFER OF ADDITIONAL CLINICAL APPLICATIONS TRAINING. THIS INFORMATION DOES NOT CONSTITUTE AN ADMISSION THAT THE DEVICE, THE COMPANY, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO A REPORTABLE EVENT.
Description of Event or Problem
BAYER MEDICAL CARE WAS NOTIFIED OF AN ALLEGED AIR INJECTION THAT HAD OCCURRED DURING A CT SCAN WHILE A PATIENT WAS CONNECTED TO A MEDRAD® CENTARGO CT INJECTION SYSTEM (SERIAL NUMBER (B)(6)). FOLLOWING THE INJECTION, THE CUSTOMER REPORTED THAT THE QUANTITY OF AIR THAT WAS VISUALIZED ON THE DISPLAYED IMAGES COULD NOT BE QUANTIFIED. THE PATIENT WAS PLACED INTO A 2-HOUR HYPERBARIC CHAMBER PROTOCOL AFTER WHICH THEY WERE PLACED ON A HIGH CONCENTRATION OXYGEN MASK AND ADMITTED TO THE HOSPITAL FOR FURTHER CARE AND TREATMENT. THE CURRENT STATUS OF THE PATIENT IS UNKNOWN.
Additional Manufacturer Narrative
THIS INVESTIGATION REMAINS IN PROGRESS. ONCE THE INVESTIGATION IS COMPLETED, A FOLLOW-UP REPORT WILL BE SUBMITTED. THE CUSTOMER CONTINUES TO USE THE MEDRAD® CENTARGO CT INJECTION SYSTEM WITH NO FURTHER ISSUES REPORTED AND HAS DECLINED THE OFFER OF ADDITIONAL CLINICAL APPLICATIONS TRAINING. THIS INFORMATION DOES NOT CONSTITUTE AN ADMISSION THAT THE DEVICE, THE COMPANY, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO A REPORTABLE EVENT.
Description of Event or Problem
BAYER MEDICAL CARE WAS NOTIFIED OF AN ALLEGED AIR INJECTION THAT HAD OCCURRED DURING A CT SCAN WHILE A PATIENT WAS CONNECTED TO A MEDRAD® CENTARGO CT INJECTION SYSTEM (B)(6). FOLLOWING THE INJECTION, THE CUSTOMER REPORTED THAT THE QUANTITY OF AIR THAT WAS VISUALIZED ON THE DISPLAYED IMAGES COULD NOT BE QUANTIFIED. THE PATIENT WAS PLACED INTO A 2-HOUR HYPERBARIC CHAMBER PROTOCOL AFTER WHICH THEY WERE PLACED ON A HIGH CONCENTRATION OXYGEN MASK AND ADMITTED TO THE HOSPITAL FOR FURTHER CARE AND TREATMENT. THE CURRENT STATUS OF THE PATIENT IS UNKNOWN.