SMARTSET GHV GENTAMICIN 40G
Received Sep 10, 2025 · Event occurred Aug 24, 2025
Report 1818910-2025-15583 · MDR key 23016433
Device
Generic name
Bone Cement : Bone Cement
Manufacturer
Depuy Orthopaedics Inc UsCatalog number
3095040
Lot number
4674653
Product problems
- Chemical Problem
- Chemical Problem
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
PRODUCT COMPLAINT #(B)(4). INVESTIGATION SUMMARY: NO DEVICE ASSOCIATED WITH THIS REPORT WAS RECEIVED FOR EXAMINATION. THE PRODUCT INVESTIGATION FOUND NO EVIDENCE SUSPECTING AN ERROR IN THE MANUFACTURING OR MATERIAL THAT WOULD BE A CONTRIBUTING FACTOR IN THE REPORTED ALLEGATION(S). A RECORDS EVALUATION (MRE) WAS NOT PERFORM. AS PART OF OUR COMPANY QUALITY SYSTEM PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND DISTRIBUTED TO APPROVED SPECIFICATIONS. ADDITIONAL COMPLAINT INFORMATION MONITORING FOR POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AS PART OF THE POST-MARKET SURVEILLANCE. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, THE INVESTIGATION WILL BE UPDATED AS APPLICABLE. DEVICE HISTORY LOT: THE PRODUCT INVESTIGATION FOUND NO EVIDENCE SUSPECTING AN ERROR IN THE MANUFACTURING OR MATERIAL THAT WOULD BE A CONTRIBUTING FACTOR IN THE REPORTED ALLEGATION(S). A MANUFACTURING RECORDS EVALUATION (MRE) WAS NOT PERFORMED.
Description of Event or Problem
IT WAS REPORTED THAT THE CEMENT WAS HEATING UP AT 13 MINUTES AND HARDENING AT 15 MINS - WHICH IS 2 MINS SLOWLY THAN THE USUAL ANTICIPATED TIMING. THERE WAS A 2 MINUTES OF SURGICAL DELAY. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THERE WAS NO PATIENT CONSEQUENCE.
Additional Manufacturer Narrative
PRODUCT COMPLAINT#: (B)(4). THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803 (AND/OR PART 4, AS APPLICABLE). THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY SYNTHES, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. ADDITIONAL NARRATIVE: G4: DEVICE IS NOT DISTRIBUTED IN THE UNITED STATES BUT IS SIMILAR TO DEVICE MARKETED IN THE USA. "DMF#: 13704, TRADE NAME: GENTAMICIN SULPHATE, ACTIVE INGREDIENT(S): GENTAMICIN SULPHATE, DOSAGE FORM: POWDER, STRENGTH: 1.0G ACTIVE IN OUR CEMENTS." IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
Additional Manufacturer Narrative
PRODUCT COMPLAINT # (B)(4). THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY SYNTHES JOINT RECONSTRUCTION, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, DEPUY SYNTHES JOINT RECONSTRUCTION, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. H11 ADDITIONAL NARRATIVE: ADDED: B5. CORRECTED: B3. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE.
Description of Event or Problem
ADDITIONAL INFORMATION RECEIVED: A. HOW WAS THE CEMENT PREPARED FOR USE? CEMVAC GUN. B. WHAT EQUIPMENT WAS USED TO PREPARE / MIX THE CEMENT? CEMVAC MIXING SYSTEM. C. WHAT WAS THE TIMING TO MIX AND THE TIME BETWEEN MIXING AND SETTING? 13 MINS. D. WHAT EQUIPMENT WAS USED TO DELIVER THE CEMENT? CEMVAC GUN. E. WHAT WAS THE STORAGE TEMPERATURE OF THE CEMENT PRIOR TO USAGE? ROOM. F. WHAT WAS THE OR TEMPERATURE DURING USE OF THE CEMENT? 18 DEG. G. CAN YOU PLEASE CONFIRM THE SURGEON'S NAME? (B)(6). H. CAN YOU PLEASE CONFIRM THE QUANTITY OF THE REPORTED DEVICE? 2.