inforMED
MalfunctionCEC

ALINITY I FREE T4 REAGENT KIT

Received Sep 10, 2025 · Event occurred Sep 1, 2025

Report 3005094123-2025-00439 · MDR key 23015651

Device

Generic name

Radioimmunoassay, Free Thyroxine

Catalog number

07P70-77

Lot number

73465UD01

Product problems

  • High Test Results
  • High Test Results

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. ALL AVAILABLE PATIENT INFORMATION WAS INCLUDED. ADDITIONAL PATIENT DETAILS ARE NOT AVAILABLE. THIS REPORT IS BEING FILED ON AN INTERNATIONAL PRODUCT, LIST NUMBER 7P70-77 THAT HAS A SIMILAR PRODUCT DISTRIBUTED IN THE US, LIST NUMBER 7P70, WITH 510K/PMA/BLA NUMBER K173122.

Description of Event or Problem

THE CUSTOMER REPORTED A FALSELY ELEVATED ALINITY I FREE T4 RESULT FOR ONE PATIENT. OTHER THYROID TESTING ON THE PATIENT WAS NORMAL.THE SAMPLE WAS REPEATED WITH LOWER RESULTS. THE FOLLOWING DATA WAS PROVIDED: INITIAL FT4 RESULT = >64 PMOL/L REPEAT RESULT = 13.6 PMOL/L. REFERENCE RANGE: 9.0-19.05 PMOL/L NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

Description of Event or Problem

THE CUSTOMER REPORTED A FALSELY ELEVATED ALINITY I FREE T4 RESULT FOR ONE PATIENT. OTHER THYROID TESTING ON THE PATIENT WAS NORMAL.THE SAMPLE WAS REPEATED WITH LOWER RESULTS. THE FOLLOWING DATA WAS PROVIDED: INITIAL FT4 RESULT = >64 PMOL/L REPEAT RESULT = 13.6 PMOL/L REFERENCE RANGE: 9.0-19.05 PMOL/L NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

Additional Manufacturer Narrative

THE COMPLAINT INVESTIGATION INCLUDED A REVIEW OF DATA AND INFORMATION PROVIDED BY THE CUSTOMER, SEARCH FOR SIMILAR COMPLAINTS, TICKET TRENDING REVIEW, DEVICE HISTORY RECORD REVIEW, AND LABELING REVIEW. ADDITIONALLY, IN-HOUSE TESTING WAS PERFORMED. DATA AND INFORMATION PROVIDED BY THE CUSTOMER WERE REVIEWED AND SUPPORT THE COMPLAINT ISSUE. REVIEW OF TRACKING AND TRENDING DATA IDENTIFIED AN INCREASE IN COMPLAINTS FOR THE ALINITY I FREE T4 ASSAY AS WELL AS AN INCREASE IN COMPLAINT ACTIVITY FOR REAGENT LOT 73465UD01. HOWEVER, TESTING WAS PERFORMED UTILIZING RETAINED KIT AND NOTED THAT ALL TESTING PASSED, INDICATING THE REAGENT LOT IS PERFORMING AS EXPECTED. THE DEVICE HISTORY RECORD REVIEW DID NOT IDENTIFY ANY NONCONFORMANCES, POTENTIAL NONCONFORMANCES OR DEVIATIONS ASSOCIATED WITH LOT NUMBER 73465UD01 AND THE COMPLAINT ISSUE. LABELING WAS REVIEWED AND ADEQUATELY ADDRESSES THE CUSTOMER¿S ISSUE. BASED ON THE INVESTIGATION, NO SYSTEMIC ISSUE OR DEFICIENCY OF THE ALINITY I FREE T4 FOR LOT 73465UD01 WAS IDENTIFIED.