inforMED
MalfunctionQVV

MISHA KNEE SYSTEM

Received Sep 9, 2025 · Event occurred Aug 11, 2025

Report 3008274656-2025-00004 · MDR key 23011837

Device

Generic name

Medial Knee Implanted Shock Absorber, Product Code: Qvv

Manufacturer

Moximed, Inc.

Catalog number

2-1002

Lot number

23092503

Product problems

  • Fracture
  • Fracture

Patient

43 YR · Male

  • Implant Pain
  • Implant Pain

Narrative

Additional Manufacturer Narrative

THE FRACTURE AT THE TIBIAL END OF THE ABSORBER WAS CONFIRMED UPON REMOVAL OF THE DEVICE. THE ROOT CAUSE IS A SUPPLIER ERROR. AN INCORRECT DRILL WAS USED RESULTING IN AN INACCURATE DRILL POINT ANGLE AND REDUCED MATERIAL THICKNESS. THIS MIGHT INCREASE THE RISK OF IMPLANT FRACTURE IN ONE LOT WHERE THE SUPPLIER ERROR OCCURRED.

Description of Event or Problem

PATIENT REPORTED MEDIAL KNEE PAIN. FRACTURE AT THE TIBIAL END OF THE ABSORBER WAS OBSERVED ON X-RAY. THE DEVICE WAS REMOVED WITHOUT DIFFICULTY.

Remedial action

  • Recall