Received Sep 9, 2025 · Event occurred Aug 5, 2025
Report 1416980-2025-04543 · MDR key 23009497
Device
Generic name
Filter, Infusion Line
Manufacturer
Baxter Healthcare CorporationModel number
1C8740Catalog number
1C8740
Lot number
DR25E30062
Product problems
- Fluid/Blood Leak
- Fluid/Blood Leak
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
IT WAS REPORTED THAT AN INTERLINK SYSTEM MINIVOLUME EXTENSION SET LEAKED FROM THE AIR FILTER TUBING. THIS WAS OBSERVED PRIOR TO USE FOR A METHYLENE INFUSION. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Additional Manufacturer Narrative
SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Additional Manufacturer Narrative
ADDITIONAL INFORMATION WAS ADDED TO D9. H3, H4 AND H6. H11: THE DEVICE WAS RECEIVED FOR EVALUATION. VISUAL INSPECTION DID NOT IDENTIFY ANY ABNORMALITIES THAT COULD HAVE CONTRIBUTED TO THE REPORTED CONDITION. FUNCTIONAL TESTING WAS PERFORMED WHICH INCLUDED PRESSURE TESTING, AND NO LEAK WAS OBSERVED IN THE AIR VENT OF THE FILTER. THE REPORTED CONDITION WAS NOT VERIFIED. A BATCH REVIEW WAS CONDUCTED AND THERE WERE NO DEVIATIONS FOUND RELATED TO THIS REPORTED CONDITION DURING THE MANUFACTURE OF THIS LOT. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.