InjuryQKA
ITIND, WITH SNARE, STERILE
Received Sep 5, 2025 · Event occurred Aug 15, 2025
Report 2429304-2025-00241 · MDR key 22978730
Device
Product problems
- Break
- Break
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
IT WAS REPORTED THAT DURING THE REMOVAL OF THE SUBJECT DEVICE; THE PHYSICIAN UTILIZED A COOK FLEXOR PARALLEL ACCESS SHEATH. DUE TO ITS DESIGN, THE SHEATH CUT THE SUTURE AT THE HUB OF THE SUBJECT DEVICE. REPORTEDLY, THE PROCEDURE WAS PROLONGED BY AT LEAST OR GREATER THAN 30 MINUTES. AS A RESULT, THE PATIENT WAS MOVED TO THE OPERATING ROOM FOR SEDATION, AND THE DEVICE WAS SUCCESSFULLY REMOVED USING A CYSTOSCOPE SHEATH AND GRASPER. THERE WERE NO REPORTS OF FURTHER PATIENT IMPACT AND HARM. ADDITIONAL INFORMATION HAS BEEN REQUESTED BY OLYMPUS, BUT NO FURTHER INFORMATION WAS RECEIVED AT THIS TIME.