inforMED
InjuryKQO

DXU 840M IRIS AUTOMATED URINE MICROSCOPY ANALYZER

Received Sep 5, 2025 · Event occurred Aug 9, 2025

Report 2122870-2025-00095 · MDR key 22977423

Device

Generic name

Dxu 840m Iris

Catalog number

C76947

Product problems

  • Leak/Splash
  • Leak/Splash

Patient

NA · Unknown

  • Fall
  • Fall

Narrative

Additional Manufacturer Narrative

THE BECKMAN FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE. THE FSE NOTED THAT THE RED CHECK VALVE IN THE FRONT LEFT OF THE INSTRUMENT WAS BROKEN; ONE OF THE TABS HOLDING THE TWO HALVES OF THE VALVE TOGETHER HAD BROKEN OFF. TO RESOLVE THE ISSUE, THE FSE REPLACED THE RED CHECK VALVE. THE OPERATOR WAS ASSESSED BY THEIR OCCUPATIONAL HEALTH DEPARTMENT THE OPERATOR HAD BRUISED BOTH KNEES, BRUISED AN ELBOW, AND REPORTED SHOULDER PAIN. THE OPERATOR WAS TAKING PAIN MEDICATION; HOWEVER, THE CUSTOMER DID NOT DIVULGE WHETHER THE PAIN MEDICATION WAS PRESCRIBED OR OVER THE COUNTER. ADDITIONALLY, THE OPERATOR WAS ICING THE AFFECTED AREAS AND RESTING. WITH THE INFORMATION AVAILABLE IT COULD NOT BE CONFIRMED WHETHER OR NOT MEDICAL INTERVENTION IN THE FORM OF PRESCRIBED MEDICATION WAS REQUIRED. THIS REPORT HAS BEEN SUBMITTED IN AN ABUNDANCE OF CAUTION. BECKMAN COULTER INTERNAL IDENTIFIER (B)(4).

Description of Event or Problem

THE CUSTOMER¿S DXU 840M IRIS INSTRUMENT HAD AN UNCONTAINED LEAK RESULTING IN AN OPERATOR SLIPPING AND FALLING TO THE FLOOR. THE OPERATOR WAS WEARING THE PROPER PERSONAL PROTECTIVE EQUIPMENT (PPE) HOWEVER THEY CAME INTO CONTACT WITH THE LEAK UPON FALLING. THE LEAK CONSISTED OF LAMINA. THERE WAS NO DIRECT CONTACT TO MUCOUS MEMBRANES OR OPEN WOUNDS. THE OPERATOR WHO HAD SLIPPEDL ON THE LIQUID ON THE FLOOR WAS EXPECTED TO REPORT TO THEIR SITE'S OCCUPATIONAL HEALTH. NO ERRONEOUS PATIENT RESULTS WERE RELEASED OUT FROM THE CUSTOMER¿S LABORATORY IN RELATION TO THIS EVENT.