inforMED
MalfunctionCDP

ELECSYS FT3 III

Received Sep 4, 2025 · Event occurred Aug 15, 2025

Report 1823260-2025-02727 · MDR key 22962681

Device

Generic name

Total Triiodothyronine Test

Manufacturer

Roche Diagnostics

Model number

FT3 III

Catalog number

09005811214

Lot number

ASKU

Product problems

  • High Test Results
  • High Test Results

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

THERE WAS AN ALLEGATION OF QUESTIONABLE THYROID RESULTS FOR ONE PATIENT FROM THE COBAS PRO SSU. THIS MEDWATCH IS FOR THE ELECSYS FT3 III ASSAY. THE MEDWATCH'S WITH A1 PATIENT IDENTIFIERS (B)(6) AND (B)(6) FOR THE OTHER INVOLVED ASSAYS.

Additional Manufacturer Narrative

THE COBAS PRO SSU SERIAL NUMBER WAS NOT PROVIDED. SAMPLE MATERIAL FROM THE PATIENT WAS REQUESTED FOR FURTHER INVESTIGATION. THE INVESTIGATION IS ONGOING.

Additional Manufacturer Narrative

TESTING OF THE PROVIDED PATIENT SAMPLE REVEALED THE PRESENCE OF A STREPTAVIDIN INTERFERING FACTOR, WHICH CAUSED THE FALSELY ELEVATED VALUES. PER PRODUCT LABELING FOR THE ASSAY: IN RARE CASES, INTERFERENCE DUE TO EXTREMELY HIGH TITERS OF ANTIBODIES TO ANALYTE-SPECIFIC ANTIBODIES, STREPTAVIDIN OR RUTHENIUM CAN OCCUR. THESE EFFECTS ARE MINIMIZED BY SUITABLE TEST DESIGN. FOR DIAGNOSTIC PURPOSES, THE RESULTS SHOULD ALWAYS BE ASSESSED IN CONJUNCTION WITH THE PATIENT¿S MEDICAL HISTORY, CLINICAL EXAMINATION AND OTHER FINDINGS.