CLINITEK STATUS+

Received Sep 2, 2025 · Event occurred Aug 4, 2025

Report 3002637618-2025-00067 · MDR key 22931576

Device

Generic name

Clinitek Status+

Catalog number

10379675

Product problems

  • Smoking
  • Smoking

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

UPON INVESTIGATION, DAMAGED POWER CORD WAS IDENTIFIED. THE DEVICE WAS FUNCTIONING WHEN TESTED WITH ANOTHER POWER CORD. THE CAUSE OF POWER CORD DAMAGE IS UNKNOWN.

Additional Manufacturer Narrative

SIEMENS REVIEWED THE INFORMATION PROVIDED BY THE CUSTOMER AND NOTED THAT THERE WERE NO ISSUES WITH THE POWER OUTLET. THE INSTRUMENT AND POWER SUPPLY WERE REQUESTED TO BE RETURNED AS WELL AS MORE INFORMATION FOR FURTHER INVESTIGATION. THE CAUSE OF THIS EVENT IS UNKNOWN.

Description of Event or Problem

CUSTOMER REPORTED THAT THE STATUS+ DEVICE PRODUCED A SMOKING SMELL. VISIBLE SMOKE WAS OBSERVED. THERE IS NO REPORT OF INJURY DUE TO THIS EVENT.