VAC PAC SIZE 30
Received Aug 29, 2025 · Event occurred Jun 16, 2025
Report 9612330-2025-00017 · MDR key 22926697
Device
Generic name
Vac Pac Size 30
Manufacturer
Natus Medical Incorporated Dba Excel-tech Ltd.Model number
51630Catalog number
51630
Lot number
N042117-08
Product problems
- Excessive Heating
- Excessive Heating
Patient
9 YR · Male
- Burn(s)
- Erythema
- Partial thickness (Second Degree) Burn
- Blister
- Burn(s)
- Erythema
- Partial thickness (Second Degree) Burn
- Blister
Narrative
Description of Event or Problem
VAC PAC SIZE 30 - THE CUSTOMER REPORTED THAT THE PATIENT ENDURED 2ND DEGREE BURNS WHILE USING THE VACPAC. THE CUSTOMER STATED THAT THE VACPAC WAS UNDER THE DRAPES AND HEATED UP, AND THAT THE HEAT SOURCE WAS LIKELY AN UNDER-BODY FLUID BED WARMING BLANKET. THE CUSTOMER STATES HE BELIEVES THE WARMING SOURCE HEATED UP THE VAC PAC UNDER THE DRAPES AND OVER SEVERAL HOURS THE VAC PAC BECAME HOT ENOUGH TO CONTRIBUTE TO THE BURN THIS PATIENT RECEIVED.
Additional Manufacturer Narrative
FOLLOW UP REPORT 001 REFERENCE NATUS COMPLAINT# (B)(4). ADDITIONAL INFORMATION ADDED TO THE BELOW SECTIONS WITH REFERENCE TO RELATED REPORT# MW5174570 RECEIVED 17 SEPT 2025. SECTION A2, SECTION A3A - PATIENT INFORMATION. SECTION B3 - DATE OF EVENT. SECTION H6 - HEALTH EFFECTS CODES. SECTION H10 - ADDED REFERECE TO RELATED REPORT NUMBER - MW5174570.
Additional Manufacturer Narrative
INITIAL REPORT REFERENCE NATUS COMPLAINT#: (B)(4). INSTALL DATE: ON (B)(6) 2017. PER DOC-040662 REV E VAC-PAC RISK ANALYSIS, HAZARD ID: 13.1, HAZARD SITUATION: VAC-PAC PAD IS NOT USED AS INTENDED, HARM: POTENTIAL INDIRECT PATIENT INJURY, CAUSE: USER DOESN'T HAVE SUFFICIENT KNOWLEDGE ABOUT THE PRODUCT, SEVERITY: LOW. THIS CUSTOMER USED THE VAC-PAC ON A PATIENT USING A WARMING SOURCE MOST LIKELY UNDER THE PATIENT FOR SEVERAL HOURS, WHICH CONTRIBUTED TO THE PATIENT RECEIVING A 2ND DEGREE BURN. THE VAC-PAC DID NOT MALFUNCTION. THE VAC PAC USER MANUAL (031817-EN REV D) STATES, "HIGH TEMPERATURES WILL DAMAGE THE VAC-PAC; NEVER STEAM AUTOCLAVE VAC-PACS" AND LISTS THE OPERATING TEMPERATURE AS "18 °C TO 29 ° CELSIUS". HISTORY OF COMPLAINTS WAS REVIEWED AND IT IS NOTED THAT THIS IS THE FIRST TIME A PATIENT RECEIVED A BURN WHILE THE VAC-PAC WAS IN USE. NO RELATED CAPAS. INVESTIGATION RESULT CODE: NEURO SBU/USER ERROR.
Description of Event or Problem
VAC PAC SIZE 30 - THE CUSTOMER REPORTED THAT THE PATIENT ENDURED 2ND DEGREE BURNS WHILE USING THE VACPAC. THE CUSTOMER STATED THAT THE VACPAC WAS UNDER THE DRAPES AND HEATED UP, AND THAT THE HEAT SOURCE WAS LIKELY AN UNDER-BODY FLUID BED WARMING BLANKET. THE CUSTOMER STATES HE BELIEVES THE WARMING SOURCE HEATED UP THE VAC PAC UNDER THE DRAPES AND OVER SEVERAL HOURS THE VAC PAC BECAME HOT ENOUGH TO CONTRIBUTE TO THE BURN THIS PATIENT RECEIVED.