inforMED
MalfunctionJGS

ATELLICA CH A-LYTE INTEGRATED MULTISENSOR (IMT NA K CL)

Received Aug 29, 2025 · Event occurred Aug 24, 2025

Report 2432235-2025-00227 · MDR key 22926509

Device

Generic name

Electrode, Ion Specific, Sodium

Model number

N/A

Catalog number

11099315

Lot number

140020

Product problems

  • Low Test Results
  • Low Test Results

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

AN OUTSIDE UNITED STATES (OUS) CUSTOMER CONTACTED THE SIEMENS HEALTHCARE DIAGNOSTICS REMOTE SERVICES CENTER (RSC) REGARDING AN ERRONEOUSLY DEPRESSED SODIUM (NA) RESULT ON ONE (1) PATIENT SAMPLE OBTAINED ON AN ATELLICA CH 930 ANALYZER. SIEMENS IS INVESTIGATING THE EVENT.

Description of Event or Problem

THE CUSTOMER REPORTED THEY OBTAINED AN ERRONEOUSLY DEPRESSED SODIUM (NA) RESULT ON ONE (1) PATIENT SAMPLE ON AN ATELLICA CH 930 ANALYZER. THE ERRONEOUSLY DEPRESSED RESULT WAS NOT REPORTED TO THE PHYSICIAN. THE SAME SAMPLE WAS REPROCESSED ON THE SAME ATELLICA CH 930 ANALYZER. A HIGHER RESULT WAS OBTAINED, CONSIDERED CORRECT, AND REPORTED TO THE PHYSICIAN. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE ERRONEOUSLY DEPRESSED SODIUM (NA) RESULT.

Additional Manufacturer Narrative

ADDITIONAL INFORMATION (11-SEP-2025): SIEMENS SERVICE OPERATIONS SUPPORT (SOS) CONCLUDED THE INVESTIGATION OF THE EVENT. QUALITY CONTROL (QC) RECOVERED WITHIN THE CUSTOMER¿S ACCEPTABLE RANGES. A CUSTOMER SERVICE ENGINEER (CSE) PERFORMED SERVICE ACTIONS ON THE ATELLICA CH 930 ANALYZER, AS PART OF TROUBLESHOOTING. NO OTHER DISCORDANT RESULTS WERE REPORTED AFTER THESE SERVICE ACTIONS. THE CAUSE OF THE EVENT IS UNKNOWN. A POTENTIAL PRODUCT ISSUE WAS NOT IDENTIFIED. THE CUSTOMER IS OPERATIONAL. THE DEVICE IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION IS REQUIRED. IN SECTION H6, THE INVESTIGATION FINDING AND INVESTIGATION CONCLUSION CODES WERE UPDATED. INITIAL MDR 2432235-2025-00227 WAS FILED ON 29-AUG-2025.