inforMED
InjuryIMK

IBOT PMD

Received Aug 29, 2025 · Event occurred Aug 2, 2025

Report 3014522447-2025-00008 · MDR key 22923766

Device

Generic name

Ibot

Model number

IBOT PMD

Catalog number

MMPX-10000-000

Lot number

N/A

Product problems

  • Use of Device Problem
  • Use of Device Problem

Patient

67 YR · Male

  • Bone Fracture(s)
  • Multiple Fractures
  • Bone Fracture(s)
  • Multiple Fractures

Narrative

Additional Manufacturer Narrative

ON (B)(6) 2025, MOBIUS MOBILITY WAS NOTIFIED BY THE USER'S CAREGIVER REGARDING A STAIR-RELATED INCIDENT RESULTING IN INJURY. THE DEVICE WAS NOT RETURNED; HOWEVER, THERE WAS ACCESS TO RETRIEVE EVENT LOGS AND BLACK BOX DATA REMOTELY ON (B)(6) 2025. ADDITIONALLY, ON (B)(6) 2025 MOBIUS MOBILITY WAS ABLE TO EVALUATE AND TEST THE DEVICE AT THE USERS HOME AND CONFIRMED THAT THERE WAS NO MALFUNCTION OF THE DEVICE. THE INVESTIGATION WAS CONDUCTED BY ON SITE TESTING OF THE IBOT INVOLVED IN THE REPORTED INCIDENT, AND REVIEWING SERVICE NOTES, DEVICE DATA AND REVIEW OF THE PRODUCTION & PATIENT RECORDS. DUE TO DATA WRAPPING AND LIMITATION OF EVENT LOG MEMORY CAPACITY, ONLY ALARM LOG DATA IS AVAILABLE FOR THE DATE INVOLVING THE REPORTED INCIDENT AT THE TIME OF ACCESSING THE DATA ON THE DEVICE. AT 13:06:34 GMT ON (B)(6) 2025 THE DEVICE REPORTED A CONTROLLER FAILURE CONDITION DUE TO PITCH LIMIT EXCEEDED. NO OTHER ALARM DATA IS PRESENT IN THE LOGS RELEVANT TO THE REPORTED INCIDENT. BASED ON THE EVALUATION/TESTING OF THE DEVICE AND THE LIMITED DATA AVAILABLE, NO DEVICE MALFUNCTION IS INDICATED IN THE REPORTED INCIDENT.

Description of Event or Problem

AN IBOT USER CALLED ON (B)(6) 2025, TO REPORT THEY HAD AN INCIDENT IN STAIR MODE RESULTING IN AN INJURY. THE IBOT OCCUPANT BROKE HIS FIBULA & TIBIA, AFTER THE DEVICE TIPPED OVER ON THE STAIRS DURING AN ASSISTED STAIR CLIMB. THE ASSISTANT ALSO FELL DURING THE EVENT RESULTING IN NO INJURIES THAT REQUIRED MEDICAL INTERVENTION.

Remedial action

  • Inspection