MalfunctionMLR

AMSCO V-PRO

Received Aug 28, 2025 · Event occurred Jul 30, 2025

Report 3005899764-2025-00071 · MDR key 22919841

Device

Generic name

V-pro Max Sterilizer

Model number

V-PRO MAX

Catalog number

VP30002101

Product problems

  • Use of Device Problem
  • Use of Device Problem

Patient

NA · Unknown

  • Burn(s)
  • Burn(s)

Narrative

Description of Event or Problem

THE USER FACILITY REPORTED THAT AN EMPLOYEE OBTAINED A BURN AFTER HANDING INSTRUMENTS THAT WERE PROCESSED IN THEIR V-PRO MAX STERILIZER. THE EMPLOYEE SOUGHT MEDICAL TREATMENT; THE USER FACILITY DID NOT DISCLOSE IF MEDICAL TREATMENT WAS ADMINISTERED.

Additional Manufacturer Narrative

A STERIS SERVICE TECHNICIAN ARRIVED ONSITE FOLLOWING THE REPORTED EVENT TO INSPECT THE UNIT AND FOUND THE UNIT TO BE OPERATING PROPERLY. THE UNIT WAS TESTED, CONFIRMED TO BE OPERATING ACCORDING TO SPECIFICATION, AND RETURNED TO SERVICE. NO REPAIRS WERE MADE. WHILE ONSITE, STERIS WAS INFORMED THAT THE EMPLOYEE WAS NOT WEARING PROPER PPE, SPECIFICALLY GLOVES, WHILE OPERATING THE STERILIZER. THE V-PRO MAX STERILIZER OPERATOR MANUAL STATES (1-2), "DANGER - CHEMICAL INJURY HAZARD: ANY VISIBLE LIQUIDS IN THE CHAMBER OR ON THE LOAD MUST BE TREATED AS CONCENTRATED HYDROGEN PEROXIDE. OBSERVE ALL HYDROGEN PEROXIDE HANDLING PRECAUTIONS. WHEN HANDLING HYDROGEN PEROXIDE, WEAR APPROPRIATE PERSONAL PROTECTIVE EQUIPMENT."THE V-PRO MAX OPERATOR MANUAL, (A-1) FURTHER STATES, "DRY ALL ITEMS THOROUGHLY. ENSURE ALL MOISTURE IS REMOVED FROM ALL INTERNAL PARTS (INCLUDING LUMENS). IF NOT, RESIDUAL HYDROGEN PEROXIDE MAY REMAIN AT CYCLE COMPLETION AND/OR A CYCLE ABORT OCCURS. ONLY DRY ITEMS ARE TO BE PLACED IN STERILIZATION UNIT." WHILE ONSITE THE TECHNICIAN COUNSELED USER FACILITY PERSONNEL ON THE IMPORTANCE OF WEARING PROPER PPE, SPECIFICALLY GLOVES, WHILE OPERATING THEIR V-PRO MAX STERILIZER AND PROPERLY DRYING INSTRUMENTS PRIOR TO PROCESSING. NO ADDITIONAL ISSUES HAVE BEEN REPORTED.