ALINITY I FREE T4 REAGENT KIT
Received Aug 28, 2025 · Event occurred Jun 20, 2025
Report 3005094123-2025-00420 · MDR key 22910982
Device
Generic name
Radioimmunoassay, Free Thyroxine
Manufacturer
Abbott Ireland Diagnostics DivisionCatalog number
07P70-30
Lot number
70400UD00
Product problems
- High Test Results
- High Test Results
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
THE CUSTOMER OBSERVED FALSELY ELEVATED ALINITY I FREE T4 FOR ONE PATIENT. THE FOLLOWING RESULTS WERE PROVIDED (REFERENCE RANGE 9.0 - 19.0 PMOL/L): ON (B)(6) 2025, SID (B)(6), INITIAL FREE T4 RESULT (B)(6)= 28.7 PMOL/L; (B)(6) 2024, TSH RESULT= 0.991; FT4 RESULT= 12.9 PMOL/L . THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
Additional Manufacturer Narrative
COMPLETE ENTRY FOR SECTION A1 - PATIENT IDENTIFIER: (B)(6), COMPLETE ENTRY FOR SECTION E1 - PHONE NUMBER: (B)(6). AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. ALL AVAILABLE PATIENT INFORMATION WAS INCLUDED. ADDITIONAL PATIENT DETAILS ARE NOT AVAILABLE.
Description of Event or Problem
THE CUSTOMER OBSERVED FALSELY ELEVATED ALINITY I FREE T4 FOR ONE PATIENT. THE FOLLOWING RESULTS WERE PROVIDED (REFERENCE RANGE 9.0 - 19.0 PMOL/L): ON (B)(6) 2025, SID (B)(6), INITIAL FREE T4 RESULT (B)(6) = 28.7 PMOL/L; (B)(6) 2024, TSH RESULT= 0.991; FT4 RESULT= 12.9 PMOL/L. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
Additional Manufacturer Narrative
THE COMPLAINT INVESTIGATION INCLUDED A REVIEW OF DATA AND INFORMATION PROVIDED BY THE CUSTOMER, SEARCH FOR SIMILAR COMPLAINTS, TICKET TRENDING REVIEW, DEVICE HISTORY RECORD REVIEW, IN HOUSE TESTING, AND LABELING REVIEW. DATA AND INFORMATION PROVIDED BY THE CUSTOMER WERE REVIEWED AND SUPPORT THE COMPLAINT ISSUE. REVIEW OF TRACKING AND TRENDING DATA FOR THE ALINITY I FREE T4 ASSAY DID IDENTIFY AN INCREASE IN COMPLAINTS. ADDITIONALLY, AN INCREASE IN COMPLAINT ACTIVITY WAS IDENTIFIED FOR REAGENT LOT 70400UD00. HOWEVER, IN HOUSE TESTING WAS PERFORMED UTILIZING RETAINED KITS OF LOT 70400UD00. ALL TESTING MET ACCEPTANCE CRITERIA AND THE PRODUCT IS PERFORMING AS EXPECTED. A DEVICE HISTORY RECORD REVIEW DID NOT IDENTIFY ANY NONCONFORMANCES, POTENTIAL NONCONFORMANCES OR DEVIATIONS ASSOCIATED WITH THE COMPLAINT ISSUE. LABELING WAS REVIEWED WHICH ADEQUATELY ADDRESSES THE CURRENT ISSUE. BASED ON THE INVESTIGATION, NO SYSTEMIC ISSUE OR DEFICIENCY OF THE ALINITY I FREE T4 REAGENT LOT 70400UD00 WAS IDENTIFIED.