STELLANT DUAL CT INJECTOR
Received Aug 27, 2025 · Event occurred Aug 3, 2025
Report 2520313-2025-00021 · MDR key 22902134
Device
Generic name
Ct Injection System
Manufacturer
Bayer Medical Care IncModel number
85577026Catalog number
SCT FLEX
Product problems
- Adverse Event Without Identified Device or Use Problem
- Adverse Event Without Identified Device or Use Problem
Patient
73 YR · Male
- Air Embolism
- Air Embolism
Narrative
Additional Manufacturer Narrative
A SYSTEM SERVICE CHECK OF THE MEDRAD® STELLANT FLEX INJECTION SYSTEM, SERIAL NUMBER (B)(6), WAS PERFORMED ON (B)(6) 2025, WHICH CONFIRMED THAT THE INJECTOR WAS OPERATING WITHIN BAYER SPECIFICATIONS. THE DISPOSABLES THAT WERE IN USE DURING THE PROCEDURE WERE DISCARDED BY THE SITE; THEREFORE, THEY ARE NOT AVAILABLE FOR EVALUATION. THE CUSTOMER WAS UNABLE TO PROVIDE THE LOT NUMBER OF THE DISPOSABLES; THEREFORE, TESTING OF RETAINED SAMPLES WAS NOT POSSIBLE. THE OFFER OF ADDITIONAL CLINICAL APPLICATIONS TRAINING WAS DECLINED BY THE CUSTOMER. THE MEDRAD® STELLANT FLEX CT INJECTION SYSTEM OPERATION MANUAL CAUTIONS THE USER AS FOLLOWS: WARNING: AIR EMBOLISM HAZARD - SERIOUS PATIENT INJURY OR DEATH MAY RESULT. ENSURE PATIENT IS NOT CONNECTED WHILE PURGING AIR FROM SYRINGE, OR ENGAGING OR ADVANCING PLUNGER. EXPEL ALL TRAPPED AIR FROM THE SYRINGE(S), CONNECTORS, TUBING, AND CATHETER BEFORE CONNECTING THE SYSTEM TO THE PATIENT. TO MINIMIZE AIR EMBOLIZATION RISKS, ENSURE THAT ONE OPERATOR IS DESIGNATED THE RESPONSIBILITY OF FILLING THE SYRINGE(S). DO NOT CHANGE OPERATORS DURING THE PROCEDURE. IF AN OPERATOR CHANGE MUST OCCUR, ENSURE THAT THE NEW OPERATOR VERIFIES THAT THE FLUID PATH IS PURGED OF AIR. THIS INFORMATION DOES NOT CONSTITUTE AN ADMISSION THAT THE DEVICE, THE COMPANY, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO A REPORTABLE EVENT.
Description of Event or Problem
BAYER MEDICAL CARE INC. WAS INFORMED THAT A 73-YEAR-OLD MALE UNDERGOING A CT SCAN EXPERIENCED AN ALLEGED AIR INJECTION WHILE CONNECTED TO A MEDRAD® STELLANT FLEX INJECTION SYSTEM (SN (B)(6)). FOLLOWING THE INJECTION, THE CUSTOMER HAD REPORTED SEEING AIR BUBBLES ON THE DISPLAYED IMAGES WITHIN THE RIGHT VENTRICLE, THE RIGHT SUB-CLAVICULAR REGION, AND IN THE SUPERIOR VENA CAVA. THE PATIENT WAS ADMITTED TO THE HOSPITAL AND RECEIVED FOLLOW-UP IMAGING WHICH SHOWED THAT THE AIR HAD RESOLVED. THE PATIENT IS REPORTED AS DOING WELL.