InjuryFHO
ORIGIN
Received May 26, 1999 · Event occurred May 18, 1999
Report 228960 · MDR key 228960
Device
Generic name
Preperitoneal Distention Balloon
Manufacturer
Guidant/origin MedsystemModel number
*Catalog number
PDB1000
Lot number
0102991
Product problems
- Balloon rupture
- Replace
- Physical Resistance
- Device Or Device Fragments Location Unknown
Patient
37 YR
- Internal Organ Perforation
- Surgery, prolonged
- Therapy/non-surgical treatment, additional
- Hematuria
- Surgical procedure, additional
- Hospitalization required
Narrative
Description of Event or Problem
WHILE HAVING A LAPAROSCOPIC REPAIR OF RT INGUINAL HERNIA, A STRUCTURAL BALLOON WAS INSERTED. BALLOON INFLATED WITH 4 HAND SQUEEZES OF HAND INSUFFLATOR. BALLOON POPPED, SURGEON INSERTED TELESCOPE AND CONFIRMED DEFLATED BALLOON. REMOVED STRUCTURAL BALLOON & INSERTED A PRE-PERITONEAL BALLOON, AND THEN TELESCOPE & FOUND URINARY BLADDER HAD BEEN COMPROMISED. BLADDER REPAIR DONE, TEMPORARY SUPRAPUBIC TUBE IN PLACE FOR 7-14 DAYS.