BD VERITOR¿ SYSTEM FOR RAPID DETECTION OF FLU A+B CLIA-WAVED KIT
Received Aug 26, 2025 · Event occurred Jul 25, 2023
Report 3006948883-2025-00541 · MDR key 22894416
Device
Generic name
Antigens, Cf (including Cf Control), Influenza Virus A, B, C
Manufacturer
Bd Rapis Diagnostics (suzhou) Co. Ltd.Catalog number
256045
Lot number
3142973
Product problems
- Device Contamination with Chemical or Other Material
- Device Contamination with Chemical or Other Material
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
INVESTIGATION SUMMARY: THIS STATEMENT IS TO SUMMARIZE THE INVESTIGATION RESULTS REGARDING THE COMPLAINT THAT INVOLVING KIT FLU A+B 30 TEST PHYSICIAN VERITOR. THE CUSTOMER REPORTED THAT THE BLACK SPOTS WERE FOUND WHEN REAGENTS WERE DROPPED IN THE READING AREA, WHICH CAUSING ABNORMAL READINGS. BD QUALITY PERFORMS A SYSTEMATIC APPROACH TO INVESTIGATE ALL COMPLAINTS. THIS APPROACH INVOLVES REVIEW OF MANUFACTURING BATCH HISTORY RECORDS, TESTING OF RETENTION SAMPLES, AND TESTING OF CUSTOMER RETURNED SAMPLES, IF APPLICABLE. BHR (BATCH HISTORY REVIEW) ANALYSIS AND RETAIN SAMPLE TESTING WERE PERFORMED ON THE BATCH NUMBER PROVIDED. THE RESULTS WERE ACCEPTABLE, AND NO RELEVANT ISSUES WERE FOUND. THE CUSTOMER HAS PROVIDED A PHOTOGRAPH CONFIRMING BLACK SPOT IN THE TEST WINDOW OF CARTRIDGE. ADDITIONALLY, ANOTHER PHOTOGRAPH OF VERITOR FLU TEST RESULT TAKEN WITH A VERITOR READER WAS PROVIDED BY THE CUSTOMER. THE TEST RESULT INDICATED A POSITIVE FOR FLU A, BUT THE PCR RESULT WAS NEGATIVE. THIS COMPLAINT WAS CONFIRMED BASED ON THE PHOTOGRAPHIC EVIDENCE. THE ROOT CAUSE COULD NOT BE DETERMINED. A TREND ANALYSIS FOR FALSE POSITIVE AND CONTAMINATION / FOREIGN MATTER WAS CONDUCTED, NO ADVERSE TREND WAS IDENTIFIED. BD QUALITY WILL CONTINUE TO MONITOR FOR TRENDS. THERE WERE NO CORRECTIVE ACTIONS TAKEN AT THIS TIME. THIS MDR IS BEING SUBMITTED AS PART OF A RETROSPECTIVE REVIEW OF RECORDS DATING APRIL 2023-2025, UNDER CAPA: 11910483.
Description of Event or Problem
IT WAS REPORTED WHILE USING BD VERITOR¿ SYSTEM FOR RAPID DETECTION OF FLU A+B CLIA-WAVED KIT, CONTAMINATION ON THE TEST DEVICE LED TO A FALSE POSITIVE RESULT. THERE WAS NO REPORT OF PATIENT IMPACT.