TRIAGE D-DIMER PANEL
Received Aug 25, 2025
Report 3013982035-2025-00022 · MDR key 22889708
Device
Generic name
Triage D-dimer Panel
Manufacturer
Quidel Cardiovascular Inc.Model number
98100EULot number
T15202RBN
Product problems
- Incorrect, Inadequate or Imprecise Result or Readings
- Low Test Results
- Incorrect, Inadequate or Imprecise Result or Readings
- Low Test Results
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
EVENT OCCURRED IN COLOMBIA. EVENT 2: CUSTOMER REPORTED D-DIMER DISCREPANCIES ON SAMPLE (B)(6) WHILE RUNNING A CORRELATION ON TRIAGE D-DIMER LOT T15202 (326; CUTOFF: <500 NG/ML) VS WERFEN ACLTOP 300 (538; CUTOFF: 499.9 NG/ML), FINAL DIAGNOSIS: EMBOLISM AND THROMBOSIS IN OTHER VEINS. NO AE REPORTED.
Additional Manufacturer Narrative
INVESTIGATION CONCLUSION: THE CUSTOMER'S COMPLAINT OF DISCREPANT LOW ISSUES WITH D-DIMER WAS NOT REPLICATED WITH IN-HOUSE RETAIN TESTING OF TRIAGE D-DIMER LOT T15202 WITH AN IN-HOUSE CALIBRATOR. NO ISSUES WITH D-DIMER RECOVERY WERE OBSERVED, THE LOT PERFORMED AS EXPECTED. REVIEWED THE BATCH RECORD FOR TRIAGE D-DIMER LOT T15202. THE LOT MET ALL FINAL RELEASE SPECIFICATIONS. NO ISSUES WITH D-DIMER RECOVERY WERE OBSERVED. BASED ON THE INFORMATION AVAILABLE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY AND NO CORRECTIVE ACTION IS REQUIRED. THE CASE DETAILS WERE REVIEWED ALONG WITH THE COMPLAINT HISTORY OF THIS PRODUCT FOR THE REPORTED FAILURE MODE; NO ISSUES WERE IDENTIFIED. BASED ON THE INFORMATION AVAILABLE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY AN NO CORRECTIVE ACTION IS REQUIRED.