inforMED
MalfunctionDXH

HOMECG+

Received Aug 22, 2025

Report 3012267350-2025-00012 · MDR key 22874400

Device

Generic name

Transmitters And Receivers, Electrocardiograph, Telephone

Model number

V2.0

Lot number

2021 - 168283

Product problems

  • Power Problem
  • Power Problem

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

THE PATIENT REPORTS THAT THE DEVICE TURNS ON AND READS "FACTORY RESET" THEN SHUTS BACK OFF AFTER CHANGING BATTERIES. THE PATIENT HAD NO ADVERSE EVENT OR A REACTION RESULTING FROM THIS PROBLEM.

Additional Manufacturer Narrative

THE POWER BUTTON APPEARED TO BE STUCK, WHICH RESULTED IN THE "FACTORY RESET" AND DEAD BATTERIES, WHICH CAUSED THE DEVICE TO CONSTANTLY TURN ON, RESET, AND TURN OFF.

Remedial action

  • Replace