Received Aug 22, 2025
Report 3012267350-2025-00009 · MDR key 22874366
Device
Generic name
Transmitters And Receivers, Electrocardiograph, Telephone
Manufacturer
Lohman Technologies, LlcModel number
V2.0Lot number
2021 - 168283
Product problems
- Power Problem
- Power Problem
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
THE BATTERIES FROM THE DEVICE WERE REPLACED AND TESTED. THE DEVICE WAS UPDATED TO THE LATEST FIRMWARE AND WAS FOUND TO BE FULLY FUNCTION. UPON TESTING OF OTHER DEVICES WITH THE SAME ISSUE, A MALFUNCTION WAS IDENTIFIED THAT PREVENTED THE DEVICE FROM TURNING ON WHEN THE BATTERIES WERE LOW, EVEN THOUGH THE REQUIRED VOLTAGE TO TURN THE DEVICE ON WAS MET.
Description of Event or Problem
THE PATIENT REPORTS THAT THE DEVICE POWERS DOWN DESPITE MULTIPLE BATTERY CHANGES. THE PATIENT HAD NO ADVERSE EVENT OR A REACTION RESULTING FROM THIS PROBLEM.
Remedial action
- Replace