inforMED
MalfunctionDXH

HOMECG+

Received Aug 22, 2025

Report 3012267350-2025-00009 · MDR key 22874366

Device

Generic name

Transmitters And Receivers, Electrocardiograph, Telephone

Model number

V2.0

Lot number

2021 - 168283

Product problems

  • Power Problem
  • Power Problem

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE BATTERIES FROM THE DEVICE WERE REPLACED AND TESTED. THE DEVICE WAS UPDATED TO THE LATEST FIRMWARE AND WAS FOUND TO BE FULLY FUNCTION. UPON TESTING OF OTHER DEVICES WITH THE SAME ISSUE, A MALFUNCTION WAS IDENTIFIED THAT PREVENTED THE DEVICE FROM TURNING ON WHEN THE BATTERIES WERE LOW, EVEN THOUGH THE REQUIRED VOLTAGE TO TURN THE DEVICE ON WAS MET.

Description of Event or Problem

THE PATIENT REPORTS THAT THE DEVICE POWERS DOWN DESPITE MULTIPLE BATTERY CHANGES. THE PATIENT HAD NO ADVERSE EVENT OR A REACTION RESULTING FROM THIS PROBLEM.

Remedial action

  • Replace