inforMED
MalfunctionDXH

HOMECG+

Received Aug 22, 2025

Report 3012267350-2025-00007 · MDR key 22874313

Device

Generic name

Transmitters And Receivers, Electrocardiograph, Telephone

Model number

V2.0

Lot number

2021 - 168283

Product problems

  • Failure to Power Up
  • Failure to Power Up

Patient

NA · Female

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE BATTERIES IN THE DEVICE APPEARED TO BE LOW IN VOLTAGE. REPLACING TO NEW BATTERIES RESULTED IN A FULLY FUNCTIONAL DEVICE. HOWEVER, THERE WAS AN ISSUE WITH THE FIRMWARE THAT PREVENTED THE DEVICE FROM TURNING ON WHEN THE BATTERIES WERE LOW BUT THE REQUIRED VOLTAGE TO TURN THE DEVICE ON WAS MET.

Description of Event or Problem

THE PATIENT REPORTS THAT THE DEVICE DOES NOT TURN ON, EVEN DESPITE MULTIPLE BATTERY CHANGES. THE PATIENT HAD NO ADVERSE EVENT OR A REACTION RESULTING FROM THIS PROBLEM.

Remedial action

  • Replace