HOMECG+
Received Aug 22, 2025
Report 3012267350-2025-00005 · MDR key 22874250
Device
Generic name
Transmitters And Receivers, Electrocardiograph, Telephone
Manufacturer
Lohman Technologies, LlcModel number
V2.0Lot number
20203 - 171351
Product problems
- Power Problem
- Power Problem
Patient
NA · Female
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
THE PATIENT REPORTS RECEIVING A NEW DEVICE THAT POWERS ON AND THEN INDICATES THAT THERE IS LOST CONTACT AND POWERS OFF. THE PATIENT HAD NO ADVERSE EVENT OR REACTION RESULTING FROM THIS PROBLEM.
Additional Manufacturer Narrative
THE BATTERIES IN THE DEVICE WERE REPLACED AND TESTED. THE DEVICE WAS UPDATED TO THE LATEST FIRMWARE AND WAS FOUND TO BE FULLY FUNCTIONAL. UPON TESTING OF OTHER DEVICES WITH THE SAME ISSUE A MALFUNCTION WAS IDENTIFIED THAT PREVENTED THE DEVICE FROM TURNING ON WHEN THE BATTERIES WERE LOW, EVEN THOUGH THE REQUIRED VOLTAGE TO TURN THE DEVICE ON WAS MET.
Remedial action
- Replace