Report 3012267350-2025-00001 · MDR key 22873951

Device

Generic name

Transmitters And Receivers, Electrocardiograph, Telephone

Model number

V2.0

Lot number

2021 - 168283

Product problems

  • Application Program Problem
  • Application Program Problem

Patient

NA · Female

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE DEVICE APPEARED TO BE OPERATING ON AN OLDER FIRMWARE VERSION WHICH WAS CAUSING ISSUES WHEN ATTEMPTING TO TAKE SCANS. THE DEVICE WAS TESTED, THEN UPDATED TO LATEST FIRMWARE AND WAS FOUND TO BE FUNCTIONAL. AFTER DELIVERY OF A NEW DEVICE WITH UPDATED FIRMWARE, THE PATIENT WAS ABLE TO SUCCESSFULLY OBTAIN SCANS.

Description of Event or Problem

THE PATIENT REPORTS RECEIVING AN UNABLE TO ANALYZE RESULT THE MAJORITY OF THE TIME WHEN ATTEMPTING READINGS. WHEN REVIEWING DATA UPLOADED TO THE CLOUD, THERE ARE INCONSISTENCIES IN THE DATA AVAILABLE. THE PATIENT HAD NO ADVERSE EVENT OR A REACTION RESULTING FROM THIS PROBLEM.

Remedial action

  • Replace