SIROS D-VIAL PREP SET
Received Aug 22, 2025 · Event occurred Dec 14, 2023
Report 3005579300-2025-00009 · MDR key 22873293
Device
Generic name
System, Applicator, Radionuclide, Manual
Manufacturer
Sirtex Medical Pty LimitedModel number
SIR-10200Lot number
1109047
Product problems
- Leak/Splash
- Leak/Splash
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
AT THE VERY END OF THE PROCEDURE, WHILE FLUSHING THROUGH THE D-LINE A FEW DROPLETS OF D5W WERE NOTICED ESCAPING THROUGH THE SEPTUM OF THE D-VIAL AND ROLLING OVER THE EDGE. NOTHING ESCAPED THE SIROS SYSTEM AND AFTER SURVEYING AND CONDUCTING THE POST PROCEDURE MEASUREMENTS THERE WAS ALMOST NO RESIDUAL AND OVER 100% OF THE WRITTEN DIRECTIVE WAS DELIVERED. WHATEVER MALFUNCTION OCCURRED IT DIDN'T HAPPEN UNTIL WE WERE DONE DELIVERING THE Y90 SPHERES BUT HAD THIS HAPPENED EARLIER IN THE PROCEDURE IT MAY HAVE BEEN A DIFFERENT STORY. I TOOK A SHORT VIDEO OF THE INCIDENT TO BETTER EXPLAIN WHAT HAPPENED AND CAN PROVIDE IT IF NECESSARY.
Additional Manufacturer Narrative
THERE WAS NO REPORTED DEATH OR SERIOUS INJURY TO THE PATIENT AND THEY RECEIVED THEIR FULL DOSE. THE DEVICE WAS RETURNED TO SIRTEX FOR INVESTIGATION AND THE EVENT WAS NOT REPLICATED. A BATCH RECORD REVIEW WAS PERFORMED AND NO DISCREPANCIES WERE FOUND. A RETROSPECTIVE MDR HAS BEEN FILED OUT OF AN ABUNDANCE OF CAUTION DUE TO THE POTENTIAL FOR SERIOUS INJURY SHOULD THE FULL DOSE NOT BE ABLE TO BE ADMINISTERED.