KARLIN RETRACTOR, CRANK FRAME BODY

Received Aug 21, 2025

Report 3007208013-2025-00040 · MDR key 22864595

Device

Generic name

Karlin Retractor

Model number

46-3191

Lot number

2235193 SO416533

Product problems

  • Mechanical Jam
  • Mechanical Jam

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE DEVICE HAS BEEN RETURNED AND IS IN PROCESS OF BEING EVALUATED. THE INVESTIGATION IS ONGOING. ONCE WE HAVE ADDITIONAL RELEVANT INFORMATION IT WILL BE SUBMITTED IN A SUPPLEMENTAL REPORT.

Description of Event or Problem

COMPLAINANT ALLEGES THAT THE DEVICE IS NOT WORKING PROPERLY, IT IS NOT CRANKING. THIS OCCURRED DURING A PATIENT PROCEDURE.

Additional Manufacturer Narrative

THE DEVICE WAS RETURNED FOR EVALUATION. IT WAS CONFIRMED THAT THE DEVICE WAS NOT FUNCTIONING AS INTENDED. THE RETRACTOR WAS CLEANED, STERILIZED, AND LUBRICATED AFTER WHICH IT FUNCTIONED CORRECTLY AS INTENDED. THE ROOT CAUSE WAS INSUFFICIENT MAINTENANCE, ESPECIALLY A LACK OF OR MINIMAL LUBRICATION AS OUTLINED IN THE DEVICES IFU. IF ANY ADDITIONAL RELEVANT INFORMATION IS IDENTIFIED, IT WILL BE SUBMITTED IN A SUPPLEMENTAL REPORT.