Received Aug 21, 2025 · Event occurred Mar 12, 2025

Report 3006418479-2025-00001 · MDR key 22861152

Device

Generic name

Magnetic Stimulator

Manufacturer

Neurosoft Ltd

Model number

NS058201.014-22

Product problems

  • Off-Label Use
  • Off-Label Use

Patient

NA · Female

  • Convulsion/Seizure
  • Convulsion/Seizure

Narrative

Additional Manufacturer Narrative

IN SPITE OF THE INDICATIONS FOR USE CLEARLY STATED IN THE USER MANUAL (PAGE 6, TM058.05.003.001) FOR THE DEVICE AND THE WARNING DISPLAYED BY THE SOFTWARE REGARDING THE PROHIBITION OF THE GENERATION OF THE NON-FDA CLEARED TMS PROTOCOLS, THE CUSTOMER DECIDED TO CREATE A NEW TMS PROTOCOL CALLED "PTSD, ANXIETY, MEMORY" OFFERED BY HER FRIEND. THE PROTOCOL GENERATED BY THE CUSTOMER AND FOUND ON THE RETURNED LAPTOP, WAS ANALYZED BY THE MANUFACTURER. IT WAS CONCLUDED THAT TMS PROTOCOL FOR HIGH-FREQUENCY (10-20 HZ) MULTI-SITE STIMULATION AT 50-120% OF MOTOR THRESHOLD WAS APPLIED. THIS EXPERIMENTAL TREATMENT IS NOT CLEARED BY FDA AND NOT ALLOWED FOR USE BY THE MANUFACTURER. IT ALSO DOES NOT COMPLY WITH THE RECOMMENDATIONS OR GUIDELINES SUGGESTED BY THE CLINICAL TMS SOCIETY AND STATE-OF-ART. MOREOVER, THE PROPER WORKFLOW PRESUMES THAT TREATMENT SESSION IS PERFORMED IN THE SITTING POSITION (PROPER USE SECTION, PAGE 49 OF THE USER MANUAL) AND UNDER THE SUPERVISION OF THE MEDICAL STAFF (WARNING AND PRECAUTION SECTION OF THE USER MANUAL, PAGE 9) WHICH ALSO WAS IGNORED BY THE CUSTOMER. THE CUSTOMER DISCHARGE WAS NOT CONFIRMED BY THE VIDEO OR DURING HER EMERGENCY ROOM VISIT. AS FAR AS THE UNKNOWN, EXPERIMENTAL, NOT FDA-CLEARED AND NOT ALLOWED BY THE MANUFACTURER TMS PROTOCOL WAS USED, NEUROSOFT CONCLUDED THAT IT IS AN OFF-LABEL DEVICE USE.

Description of Event or Problem

ON (B)(6) 2025, THE CUSTOMER CONTACTED THE AUTHORIZED AGENT'S OFFICE TO TELL THAT SHE EXPERIENCED A SEIZURE WHILE TREATING HERSELF WITH THE MACHINE, USING AN UNKNOWN PROTOCOL THAT, ACCORDING TO HER, WAS DESIGNED BY ONE OF HER FRIENDS. THE CUSTOMER SENT A VIDEO OF THAT INCIDENT, THE VIDEO BEING DATED (B)(6) 2025, SHOWING SHE WAS ADMINISTERING THAT TREATMENT TO HERSELF, WITH NO ONE SEEN IN THE ROOM WITH HER. SHE IS SEEN STANDING UP WITH THE COIL ON HER HEAD AND THEN FALLING DOWN AND GOING OUT OF CAMERA VIEW, SO THERE WAS NO CONFIRMATION BY VIDEO RECORDING, OR BY ANY OBSERVER OF WHAT HAPPENED DURING OR AFTER THE FALL. ON (B)(6) 2025, THE CUSTOMER EMAILED THE AUTHORIZED AGENT TELLING THAT SHE HAD BEEN TAKEN TO THE EMERGENCY ROOM AFTER THE FALL, AND WAS DISCHARGED WITH "SEIZURE AND CONCUSSION PRECAUTIONS", BUT THERE WAS NO MENTION IN HER EMAIL OF ANY EXAM FOR SEIZURES OR EEG RECORDING DURING HER EMERGENCY ROOM VISIT. THE CUSTOMER RETURNED THE MACHINE AND ITS LAPTOP, WHERE IT WAS WITNESSED THAT SHE HAD INSTALLED A TMS PROTOCOL THAT WAS NOT INCLUDED IN HER LAPTOP PRIOR TO SHIPMENT. ONCE THE MACHINE WAS BACK, IT WAS COMPLETELY CHECKED, ITS FUNCTIONING WAS TESTED AND IT WAS FOUND THAT THE DEVICE IS IN GOOD WORKING ORDER.