REWALK 7
Received Aug 19, 2025 · Event occurred Jul 17, 2025
Report 3007615665-2025-00001 · MDR key 22836808
Device
Generic name
Powered Exoskeleton
Manufacturer
Lifeward Ltd.Model number
REWALK 7Catalog number
50-20-0005
Product problems
- Adverse Event Without Identified Device or Use Problem
- Adverse Event Without Identified Device or Use Problem
Patient
55 YR · Female
- Thrombosis/Thrombus
- Thrombosis/Thrombus
Narrative
Additional Manufacturer Narrative
FINAL REPORT AND INVESTIGATION MAIN FINDINGS: DATE: OCTOBER 20, 2025 1. THE INVESTIGATION ACTIVITIES INCLUDED THE FOLLOWING: REVIEWING DEVICE RECORDS. INTERVIEWING THE CLINICAL TRAINER AND THE PATIENT. ANALYSING THE TIMELINE OF THE PATIENT'S DEVICE USAGE PRIOR TO THE REPORTED INCIDENT. REVIEWING AVAILABLE PATIENT DATA AND PARAMETERS. CONDUCTING A DEVICE INSPECTION BY A LIFEWARD FIELD SERVICE ENGINEER, WHO FOUND NO EVIDENCE OF MALFUNCTION. EVALUATING MEDICAL INFORMATION PROVIDED BY THE PATIENT BY AN EXTERNAL MEDICAL EXPERT (PHYSICIAN) TO IDENTIFY THE POTENTIAL ROOT CAUSE. MEDICAL ASSESSMENT CONCLUSION: THE PATIENT HAD A RECURRENCE OF PRE-EXISTING PROXIMAL LOWER LIMB VENOUS THROMBOTIC DISEASE WITH OCCLUSION OF PREVIOUSLY IMPLANTED VENOUS STENTS. THE REWALK DEVICE DOES NOT IN ANY WAY INTERFERE WITH THE FUNCTION OF THE VASCULAR SYSTEM OR IMPLANTED VASCULAR STENTS. REWALK DEVICE USE WAS NOT THE CAUSE OF THE PATIENT'S VASCULAR DISEASE OR VASCULAR DISEASE COMPLICATIONS. BASED ON THE COLLECTED INFORMATION, THE CAUSE IS LIKELY RELATED TO THE USER'S PRE-EXISTING VASCULAR CONDITION AND NOT TO THE REWALK DEVICE.
Description of Event or Problem
THE FOLLOWING DATA WAS REPORTED TO THE LIFEWARD CLINICAL TRAINING MANAGER BY A PHYSICAL THERAPIST AT THE USER'S TRAINING CLINIC: ON (B)(6) 2025: THE USER REPORTED A PINCHING SENSATION IN HER RIGHT HIP DURING USE OF THE DEVICE. LATER THAT SAME DAY, DURING A FOLLOW-UP SESSION, SHE REPORTED THAT THE PINCHING DID NOT RECUR. ON (B)(6) 2025: THE USER CANCELLED HER SESSION DUE TO SWELLING IN HER LEFT LEG. THE USER BELIEVES THIS MAY BE RELATED TO THE STENTS IN HER GROIN AND POSSIBLY THE CAUSE OF THE EARLIER PINCHING SENSATION. ON (B)(6) 2025: THE USER VISITED THE ER AND WAS DIAGNOSED WITH BLOOD CLOTS IN HER LEG. THE USER IS AWAITING CONSULTATION WITH HER PHYSICIAN TO DETERMINE THE NEXT STEPS IN TREATMENT. THE INCIDENT WAS RECORDED UNDER CUSTOMER COMPLAINT NUMBER: (B)(4).
Additional Manufacturer Narrative
THE INCIDENT IS CURRENTLY UNDER INVESTIGATION. BASED ON THE INFORMATION COLLECTED SO FAR, THE CAUSE IS LIKELY RELATED TO THE USER'S PRE-EXISTING VASCULAR CONDITION AND NOT TO THE REWALK DEVICE.
Description of Event or Problem
THE FOLLOWING DATA WAS REPORTED TO THE LIFEWARD CLINICAL TRAINING MANAGER BY A PHYSICAL THERAPIST AT THE USER'S TRAINING CLINIC: (B)(6), 2025: THE USER REPORTED A PINCHING SENSATION IN HER RIGHT HIP DURING USE OF THE DEVICE. LATER THAT SAME DAY, DURING A FOLLOW-UP SESSION, SHE REPORTED THAT THE PINCHING DID NOT RECUR. (B)(6), 2025: THE USER CANCELLED HER SESSION DUE TO SWELLING IN HER LEFT LEG. THE USER BELIEVES THIS MAY BE RELATED TO THE STENTS IN HER GROIN AND POSSIBLY THE CAUSE OF THE EARLIER PINCHING SENSATION. (B)(6), 2025: THE USER VISITED THE ER AND WAS DIAGNOSED WITH BLOOD CLOTS IN HER LEG. THE USER IS AWAITING CONSULTATION WITH HER PHYSICIAN TO DETERMINE THE NEXT STEPS IN TREATMENT. STATUS: THE USER IS ON HOLD FROM TRAINING UNTIL THE MEDICAL ISSUE IS FULLY RESOLVED AND CLEARED BY HER HEALTHCARE PROVIDER. THE INCIDENT WAS RECORDED UNDER CUSTOMER COMPLAINT NUMBER: (B)(4).