inforMED
MalfunctionJFP

ROCHE 9180 ELECTROLYTE ANALYZER

Received Aug 15, 2025 · Event occurred Jul 24, 2025

Report 1823260-2025-02518 · MDR key 22817663

Device

Generic name

Electrolyte Analyzer

Manufacturer

Roche Diagnostics

Catalog number

03157334001

Product problems

  • Incorrect, Inadequate or Imprecise Result or Readings
  • Incorrect, Inadequate or Imprecise Result or Readings

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE ROOT CAUSE OF THE ISSUE WAS RELATED TO A REFERENCE ELECTRODE MANUFACTURED BY DIAMOND DIAGNOSTICS (DD REF ELECTRODE) USED WITH THE 9180 ELECTROLYTE ANALYZER. THE REFERENCE ELECTRODE USED (DD REF ELECTRODE) IS COVERED BY A ROCHE INITIATED RECALL. CUSTOMERS USING THE DD REF ELECTRODE WERE INSTRUCTED TO IMMEDIATELY STOP USING SODIUM (NA) RESULTS FROM THEIR 9180 ANALYZER. THE AFFECTED REFERENCE ELECTRODE WAS NOT DISTRIBUTED IN THE UNITED STATES. CUSTOMERS IN THE UNITED STATES USE ROCHE REFERENCE ELECTRODE AND REFERENCE ELECTRODE HOUSING. NO FURTHER ACTION IS NEEDED FOR CUSTOMERS IN THE UNITED STATES.

Description of Event or Problem

THE INITIAL REPORTER COMPLAINED OF LOW RESULTS FOR MULTIPLE PATIENT SAMPLES TESTED WITH A ROCHE 9180 SODIUM ELECTRODE (NA) ON A ROCHE 9180 ELECTROLYTE ANALYZER. A DISCREPANT LOW RESULT WAS PROVIDED FOR 1 PATIENT SAMPLE TESTED USING A NEW NA ELECTRODE AND A PREVIOUS NA ELECTRODE. THE RESULT USING THE PREVIOUS NA ELECTRODE WAS 131 MMOL/L. THE RESULT USING THE NEW NA ELECTRODE WAS 116 MMOL/L. THE RESULTS USING THE NEW NA ELECTRODE ARE IN QUESTION.

Additional Manufacturer Narrative

THE NA ELECTRODE LOT NUMBER IN QUESTION WAS 31245147 WITH AN EXPIRATION DATE OF 15-AUG-2025.