inforMED
MalfunctionBSP

SPINAL NEEDLE 18GA 3.00 IN

Received Aug 15, 2025 · Event occurred Jul 1, 2025

Report 3003152976-2025-00449 · MDR key 22814755

Device

Generic name

Needle, Conduction, Anesthetic (w/wo Introducer)

Manufacturer

Becton Dickinson

Catalog number

405247

Lot number

2010009

Product problems

  • Contamination
  • Contamination

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

ADDITIONAL INFORMATION: THE GREEN PARTICLE WAS OBSERVED IN ITS FINAL PACKAGING (TRANSFER BAG), TOGETHER WITH THE ACTIVE SUBSTANCE AND ITS EXCIPIENTS.

Additional Manufacturer Narrative

(B)(4) FOLLOW UP MDR FOR CORRECTION. CORRECTION: THIS COMPLAINT WAS DETERMINED TO NOT BE A REPORTABLE EVENT AFTER THE MDR WAS SUBMITTED. THE GREEN PARTICLE WAS NOTED WITHIN THE BAG, NOT THE FLUID PATHWAY.

Additional Manufacturer Narrative

H.3. IF A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem

IT IS REPORTED FOREIGN MATTER. EVENT DESCRIPTION: IDENTIFICATION OF A PARTICLE IN THE FINISHED PRODUCT. I AM CONTACTING YOU AS PART OF A QUALITY INVESTIGATION. DURING THE INSPECTION OF OUR LATEST PRODUCTION BATCH (DISTRIBUTED IN BAGS), WE NOTICED THE PRESENCE OF A GREEN PARTICLE MEASURING 2.24 MM. THIS PARTICLE WAS IDENTIFIED BY IR AND IS COMPOSED OF POLYCARBONATE.