DRIVEMEDICAL QUAD CANE TIP
Received Aug 14, 2025
Report 2438477-2025-00038 · MDR key 22804336
Device
Generic name
Tips And Pads, Cane, Crutch And Walker
Manufacturer
Combo (zhong Shan) Medical Equipment Co.,ltd.Model number
RTL10353Product problems
- Material Integrity Problem
- Material Integrity Problem
Patient
NA · Unknown
- Hip Fracture
- Hip Fracture
Narrative
Additional Manufacturer Narrative
THE QUAD CANE TIP WAS RETURNED TO DRIVE DEVILBISS HEALTHCARE FOR EVALUATION. THE PRODUCT WAS IN GOOD CONDITION IN ITS ORIGINAL PACKAGING. THE CANE TIP WAS REVIEWED ON A HARDWOOD FLOOR CONFIRMING THAT THERE IS GRIP WHEN LOAD IS APPLIED. NO DEFECTS WERE FOUND. COMPLAINTS DATA WAS REVIEWED FOR FURTHER INSIGHT ON THE CANE TIP DESIGN CUSTOMER CONCERN OF THE CANE TIP NOT GRIPPING THE FLOOR SURFACE. THIS COMPLAINT WAS THE ONLY ONE IDENTIFIED IN A TWO-YEAR TIME PERIOD. THE SUPPLIER WAS NOTIFIED OF THE COMPLAINT. THIS APPEARS TO BE AN ISOLATED INCIDENT FROM THE COMPLAINTS DATA REVIEW. COMPLAINTS WILL CONTINUE TO BE MONITORED FOR TRENDS.
Description of Event or Problem
DRIVE DEVILBISS HEALTHCARE WAS NOTIFIED OF AN INCIDENT INVOLVING A CANE TIP BY THE PROVIDER, WHO STATED THAT DURING USE ON A HARDWOOD FLOOR OF THEIR HOME, THE END USER FELL AND BROKE HER HIP, REQUIRING HOSPITALIZATION. THE PROVIDER REPORTED THAT THE BOTTOM OF THE CANE TIP DOES NOT GRIP TO THE HARD FLOOR SURFACE DURING USE. AS PART OF ITS COMPLAINT REVIEW AND EVALUATION PROCESS, DRIVE DEVILBISS HEALTHCARE WILL ALSO NOTIFY THE MANUFACTURER OF THE PRODUCT ABOUT THIS COMPLAINT.