WM-NP2 WORKSTATION SET 6 (EU)

Received Aug 13, 2025

Report 9611174-2025-00252 · MDR key 22794856

Device

Generic name

Wm-*p2 Series Of Endoscopy Workstations

Model number

K10021617

Product problems

  • Electrical /Electronic Property Problem
  • Electrical /Electronic Property Problem

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE THE RESULTS OF THE FINAL INVESTIGATION. UPDATED FIELDS: D8, H3, H4, H6, H11. THE DEVICE WAS NOT RETURNED TO OLYMPUS FOR INSPECTION. HOWEVER, THE DEVICE WAS EVALUATED BY A FIELD SERVICE ENGINEER AND THE CUSTOMER'S MALFUNCTION WAS CONFIRMED. BASED ON THE RESULTS OF THE INVESTIGATION, THE EVENT WAS LIKELY CAUSED BY THE SHORT-CIRCUITED AND NON-FUNCTIONAL TRANSFORMER. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.

Description of Event or Problem

NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

Additional Manufacturer Narrative

THE INVESTIGATION IS ONGOING. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETED OR IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem

IT WAS REPORTED THE SUBJECT DEVICE PRESENTED A SHORT CIRCUIT EPISODE WITH THE SMELL OF BURNING DURING SETUP/INSPECTION FOR USE, BEFORE PATIENT IN ROOM. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.