SENSICA UO SYSTEM
Received Aug 13, 2025 · Event occurred Aug 8, 2025
Report 1018233-2025-06741 · MDR key 22794605
Device
Generic name
Sensica
Manufacturer
C.r. Bard Inc. (covington) -1018233Catalog number
SCCS1001
Lot number
NA
Product problems
- Volume Accuracy Problem
- Volume Accuracy Problem
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
THE INVESTIGATION IS STILL IN PROGRESS. ONCE THE INVESTIGATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. UDI FORMAT UPDATED WITH AVAILABLE PRODUCT INFORMATION PER GUDID. H11: SECTION A THROUGH F - THE INFORMATION PROVIDED BY BD REPRESENTS ALL THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT / REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO BD.
Description of Event or Problem
IT WAS REPORTED THAT THE PATIENT WAS MOBILIZED TO A CHAIR SHORTLY BEFORE SHIFT CHANGE AS WAS TYPICAL FOR THE SENSICA URINE OUTPUT UNIT, AND WHEN THEY WENT IN FOR OUR SHIFT CHANGE DRIP CHECK, THEY COULD SEE THAT THE HEIGHT OF THE FOLEY BAG WAS JUST ABOUT LEVEL WITH MY PATIENT'S BLADDER LEVEL WHEN SEATED IN THE CHAIR. THERE WAS A SMALL AMOUNT OF URINE IN THE LINE, BUT NONE IN THE CHAMBER. WHEN THEY WENT IN AT 8 TO GET HOURLY MEASUREMENTS, THEY NOTICED THAT NONE OF THE URINE WAS ABLE TO MAKE IT TO THE BAG. SO REMOVED IT TO ALLOW THE URINE TO FLOW INTO THE CHAMBER AND CHARTED THE REMOVAL AND MY REASONING.
Description of Event or Problem
IT WAS REPORTED THAT THE PATIENT WAS MOBILIZED TO A CHAIR SHORTLY BEFORE SHIFT CHANGE AS WAS TYPICAL FOR THE SENSICA URINE OUTPUT UNIT, AND WHEN THEY WENT IN FOR OUR SHIFT CHANGE DRIP CHECK, THEY COULD SEE THAT THE HEIGHT OF THE FOLEY BAG WAS JUST ABOUT LEVEL WITH MY PATIENT'S BLADDER LEVEL WHEN SEATED IN THE CHAIR. THERE WAS A SMALL AMOUNT OF URINE IN THE LINE, BUT NONE IN THE CHAMBER. WHEN THEY WENT IN AT 8 TO GET HOURLY MEASUREMENTS, THEY NOTICED THAT NONE OF THE URINE WAS ABLE TO MAKE IT TO THE BAG. SO REMOVED IT TO ALLOW THE URINE TO FLOW INTO THE CHAMBER AND CHARTED THE REMOVAL AND MY REASONING. PER FOLLOW UP INFORMATION RECEIVED VIA TASK ON 02SEP2025, EMAIL RESPONSE RECEIVED NOTING THAT THEY BELIEVE THE ISSUE WAS CAUSED BY A DESIGN FLAW IN THE SENSICA DEVICES AND THE DEVICE WOULD NOT BE RETURNING FOR EVALUATION.
Additional Manufacturer Narrative
THE REPORTED ISSUE WAS INCONCLUSIVE. ROOT CAUSE WAS NOT ABLE TO BE ISOLATED AS UNIT WAS NOT EVALUATED. THE PATIENT WAS MOBILIZED TO A CHAIR SHORTLY BEFORE SHIFT CHANGE AS WAS TYPICAL FOR THE SENSICA URINE OUTPUT UNIT, AND WHEN THEY WENT IN FOR OUR SHIFT CHANGE DRIP CHECK, THEY COULD SEE THAT THE HEIGHT OF THE FOLEY BAG WAS JUST ABOUT LEVEL WITH MY PATIENT'S BLADDER LEVEL WHEN SEATED IN THE CHAIR. WHEN THEY WENT IN AT 8 TO GET HOURLY MEASUREMENTS, THEY NOTICED THAT NONE OF THE URINE WAS ABLE TO MAKE IT TO THE BAG. PER FOLLOW UP INFORMATION, EMAIL RESPONSE RECEIVED NOTING THAT THEY BELIEVE THE ISSUE WAS CAUSED BY A DESIGN FLAW IN THE SENSICA DEVICES AND THE DEVICE WOULD NOT BE RETURNING FOR EVALUATION. SO REMOVED IT TO ALLOW THE URINE TO FLOW INTO THE CHAMBER AND CHARTED THE REMOVAL AND MY REASONING. A REVIEW OF THE RISK DOCUMENT, (B)(4) REV 9, WAS REVIEWED FOR THIS INVESTIGATION. THE RISK DOCUMENTATION WAS ADDRESSED IN STEP #10.04. THIS OCCURRENCE DOES NOT INTRODUCE A NEW HAZARD OR HARM. IT REPRESENTS A KNOWN FAILURE MODE THAT HAS BEEN ASSESSED IN THE PRODUCT RISK MANAGEMENT FILE IN ACCORDANCE WITH APPLICABLE REGULATORY STANDARDS AND INTERNAL PROCEDURES. THE ASSOCIATED RISKS CONTINUE TO BE MONITORED AS PART OF ONGOING POST-MARKET ACTIVITIES IN COMPLIANCE WITH BOTH REGULATORY REQUIREMENTS AND LOCAL PROCEDURES. THE INSTRUCTIONS-FOR-USE ARE FOUND TO BE ADEQUATE. THE IFU CURRENTLY INSTRUCTS THE USER ON THE PROPER METHOD TO USE THIS DEVICE TO AVOID UNDUE INJURY TO THE PATIENT AND DAMAGE TO THE PRODUCT. A DHR IS NOT REQUIRED AS THE SERIAL NUMBER IS UNKNOWN. BASED ON THE RESULTS OF THE INVESTIGATION, NO ADDITIONAL ACTIONS CAN BE TAKEN. UDI FORMAT UPDATED WITH AVAILABLE PRODUCT INFORMATION PER GUDID. H11: SECTIONS A THROUGH F - THE INFORMATION PROVIDED BY BD REPRESENTS ALL THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT / REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO BD.