Received Aug 8, 2025 · Event occurred Jul 29, 2025

Report 2020813-2025-00005 · MDR key 22748705

Device

Generic name

Flowmeter

Manufacturer

Accutron, Inc.

Model number

31980

Catalog number

31980

Product problems

  • Adverse Event Without Identified Device or Use Problem
  • Adverse Event Without Identified Device or Use Problem

Patient

NA · Unknown

  • Nausea
  • Nausea

Narrative

Description of Event or Problem

DISTRIBUTOR REPORTED THAT THE UNIT IS GIVING TOO MUCH NITROUS AND MAKING PATIENTS SICK.

Additional Manufacturer Narrative

A COMPLETE AND THOROUGH TEST AND EVALUATION WAS CONDUCTED OF THIS FLOWMETER. COULD NOT DUPLICATE REPORTED PROBLEM. FLOWMETER MEETS FACTORY SPECIFICATIONS.