inforMED
MalfunctionFLE

AMSCO

Received Aug 7, 2025 · Event occurred Jul 9, 2025

Report 3005899764-2025-00067 · MDR key 22743415

Device

Generic name

Amsco 400 Steam Sterilizer

Catalog number

SR2221211011

Product problems

  • Detachment of Device or Device Component
  • Detachment of Device or Device Component

Patient

NA · Unknown

  • Bruise/Contusion
  • Bruise/Contusion

Narrative

Additional Manufacturer Narrative

A STERIS SERVICE TECHNICIAN ARRIVED ONSITE FOLLOWING THE REPORTED EVENT TO INSPECT THE UNIT AND FOUND THAT THE LATCHING SYSTEM OF THE UNIT WAS MISALIGNED. THE TECHNICIAN ADJUSTED THE PANEL SUPPORT BRACKET, TESTED THE UNIT AND CONFIRMED IT TO BE OPERATING ACCORDING TO SPECIFICATION, AND RETURNED IT TO SERVICE. NO ADDITIONAL ISSUES HAVE BEEN REPORTED.

Description of Event or Problem

THE USER FACILITY REPORTED THAT THE BOTTOM PANEL TO THEIR AMSCO 400 STERILIZER DETACHED AND CONTACTED THE EMPLOYEE'S FOOT (BRUISE). THE EMPLOYEE WENT TO THE EMERGENCY ROOM. THE USER FACILITY DID NOT DISCLOSE FURTHER MEDICAL INFORMATION.