BD BBL¿ CHROMAGAR® MRSA II

Received Aug 7, 2025 · Event occurred Jul 11, 2025

Report 1119779-2025-02560 · MDR key 22739248

Device

Generic name

Culture Media, Antimicrobial Susceptibility Test, Excluding Mueller Hinton Agar

Catalog number

215229

Lot number

5093285

Product problems

  • Incorrect, Inadequate or Imprecise Result or Readings
  • Incorrect, Inadequate or Imprecise Result or Readings

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

REPORT 5 OF 85. IT WAS REPORTED WHILE USING BD BBL¿ CHROMAGAR® MRSA II, ONE (1) PATIENT ISOLATE RESULTED AS POSITIVE FOR METHICILLIN RESISTANT STAPHYLOCOCCUS AUREUS (MRSA). UPON PERFORMING GENOME SEQUENCING OF THE ISOLATE, IT WAS CONFIRMED THE PATIENT RESULT WAS SUSCEPTIBLE TO METHICILLIN. THERE WAS NO ADVERSE IMPACT REPORTED.

Additional Manufacturer Narrative

B3. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. THE INITIAL REPORTER PROVIDED MAY 2025 WITH NO FURTHER DATE INFORMATION. E4. FDA NOTIFIED: THE INITIAL REPORTER ALSO NOTIFIED THE FDA IN JULY 2025. NO FURTHER DATE INFORMATION IS AVAILABLE. MEDSUN REPORT # (B)(4). A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative

THE FOLLOWING FIELDS HAVE BEEN UPDATED WITH CORRECTED AND/OR ADDITIONAL INFORMATION. D4. MEDICAL DEVICE LOT#: 5093285 D4. MEDICAL DEVICE EXPIRATION DATE: 23-JUNE-2025 H4. DEVICE MANUFACTURE DATE: 03-APRIL-2025 D4. UNIQUE IDENTIFIER (UDI) #: (B)(4). INVESTIGATION SUMMARY: DURING MANUFACTURING OF MATERIAL 215229, MEDIA IS FORMULATED AND SENT THROUGH A HIGH TEMPERATURE SHORT TIME STERILIZER TO REMOVE BIOBURDEN. THE PETRI DISHES ARE SUBJECTED TO UV RADIATION TO DECREASE BIOBURDEN. THE PETRI DISHES ARE FILLED IN A POSITIVE PRESSURE HEPA FILTERED ENVIRONMENT. THE FILLED PLATES ARE COOLED AND IMMEDIATELY WRAPPED INTO SLEEVES TO DECREASE THE INTRODUCTION OF MICROBES. SLEEVES ARE THEN PACKAGED INTO CARTONS AND THEN TRANSFERRED TO A REFRIGERATED TRUCK (2 TO 8 DEGREES C) FOR SHIPMENT TO THE DISTRIBUTOR. BD DISTRIBUTORS ARE PROVIDED WITH THE STORAGE GUIDELINES FOR THE SHIPPING AND HANDLING OF BD MEDIA OF 2 TO 8 DEGREES C IN A DARK PLACE. THE BATCH HISTORY RECORD FOR BATCH 5093285 WAS SATISFACTORY AND NO QUALITY NOTIFICATIONS WERE GENERATED DURING MANUFACTURING AND INSPECTION. ALL PERFORMANCE TESTING ON THIS BATCH WAS SATISFACTORY AT THE TIME OF RELEASE. ALL BATCHES ARE TESTED PRIOR TO RELEASE AND RESULTS REPORTED ON THE CERTIFICATE OF ANALYSIS WHICH CAN BE OBTAINED AT WWW.BD.COM/REGDOCS. PLATE CHROMAGAR MRSA II IS STABILITY TESTED BIENNIALLY FOR BIOLOGICAL PERFORMANCE TO ENSURE PERFORMANCE IS SATISFACTORY THROUGHOUT SHELF LIFE WITH THE ORGANISMS THAT ARE REPORTED ON THE CERTIFICATE OF ANALYSIS. PHOTOS AND RETURNS WERE NOT RECEIVED FOR THIS INVESTIGATION. RETENTION SAMPLES WERE TESTED PER THE STANDARD PERFORMANCE PROCEDURE WHICH IS ALSO DESCRIBED IN THE IFU (INSTRUCTIONS FOR USE, AVAILABLE ON BD.COM/E-LABELING). TESTING CONDUCTED FOR THE INVESTIGATION HAS REPLICATED THE PERFORMANCE DEFECT. THIS COMPLAINT IS CONFIRMED. A TREND IN PERFORMANCE COMPLAINTS HAS BEEN IDENTIFIED FOR BATCH 5093285. BD HAS INITIATED A CAPA (CORRECTIVE AND PREVENTATIVE ACTION) TO FORMALLY INVESTIGATE THE PERFORMANCE ISSUE IDENTIFIED FOR BATCH 5093285. BD WILL CONTINUE TO TREND COMPLAINTS FOR PERFORMANCE.

Description of Event or Problem

REPORT 5 OF 85. IT WAS REPORTED WHILE USING BD BBL¿ CHROMAGAR® MRSA II, ONE (1) PATIENT ISOLATE RESULTED AS POSITIVE FOR METHICILLIN RESISTANT STAPHYLOCOCCUS AUREUS (MRSA). UPON PERFORMING GENOME SEQUENCING OF THE ISOLATE, IT WAS CONFIRMED THE PATIENT RESULT WAS SUSCEPTIBLE TO METHICILLIN. THERE WAS NO ADVERSE IMPACT REPORTED.