ALINITY I FREE T4 REAGENT KIT
Received Aug 7, 2025 · Event occurred Jul 25, 2025
Report 3005094123-2025-00387 · MDR key 22736648
Device
Generic name
Radioimmunoassay, Free Thyroxine
Manufacturer
Abbott Ireland Diagnostics DivisionCatalog number
07P70-77
Lot number
73465UD01
Product problems
- High Test Results
- High Test Results
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
THE CUSTOMER REPORTED A FALSELY ELEVATED ALINITY I FREE T4 RESULT ON A PATIENT. RESULTS PROVIDED: 32.58 / 13.52 PMOL/L (REFERENCE RANGE: 9.01-19.05 PMOL/L NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Additional Manufacturer Narrative
THE COMPLAINT INVESTIGATION INCLUDED A REVIEW OF DATA AND INFORMATION PROVIDED BY THE CUSTOMER, SEARCH FOR SIMILAR COMPLAINTS, TICKET TRENDING REVIEW, DEVICE HISTORY RECORD REVIEW, AND LABELING REVIEW. ADDITIONALLY, IN-HOUSE TESTING WAS PERFORMED. DATA AND INFORMATION PROVIDED BY THE CUSTOMER WERE REVIEWED AND SUPPORT THE COMPLAINT ISSUE. REVIEW OF TRACKING AND TRENDING DATA IDENTIFIED AN INCREASE IN COMPLAINTS FOR THE ALINITY I FREE T4 ASSAY AS WELL AS AN INCREASE IN COMPLAINT ACTIVITY FOR REAGENT LOT 73465UD01. HOWEVER, TESTING WAS PERFORMED UTILIZING RETAINED KITS OF LOT 73465UD01 AND NOTED THAT ALL TESTING PASSED, INDICATING THE REAGENT LOT IS PERFORMING AS EXPECTED. THE DEVICE HISTORY RECORD REVIEW DID NOT IDENTIFY ANY NONCONFORMANCES, POTENTIAL NONCONFORMANCES OR DEVIATIONS ASSOCIATED WITH LOT NUMBER 73465UD01 AND THE COMPLAINT ISSUE. LABELING WAS REVIEWED AND ADEQUATELY ADDRESSES THE CUSTOMER¿S ISSUE. BASED ON THE INVESTIGATION, NO SYSTEMIC ISSUE OR DEFICIENCY OF THE ALINITY I FREE T4 FOR LOT 73465UD01 WAS IDENTIFIED.
Additional Manufacturer Narrative
AN EVALUATION IS IN PROCESS. A FOLLOW UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. ALL AVAILABLE PATIENT INFORMATION WAS INCLUDED. ADDITIONAL PATIENT DETAILS ARE NOT AVAILABLE.
Description of Event or Problem
THE CUSTOMER REPORTED A FALSELY ELEVATED ALINITY I FREE T4 RESULT ON A PATIENT. RESULTS PROVIDED: 32.58 / 13.52 PMOL/L (REFERENCE RANGE: 9.01-19.05 PMOL/L. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.