ANASTOCLIP GC CLOSURE SYSTEM - VASCULAR & NEURO
Received Aug 6, 2025 · Event occurred Jul 7, 2025
Report 1220948-2025-00130 · MDR key 22729546
Device
Generic name
Closure System
Manufacturer
Lemaitre Vascular, Inc.Catalog number
4008-06
Lot number
AGC00002562
Product problems
- Failure to Form Staple
- Mechanical Jam
- Failure to Form Staple
- Mechanical Jam
Patient
NA · Unknown
- Infiltration into Tissue
- Infiltration into Tissue
Narrative
Description of Event or Problem
IT WAS REPORTED THAT DURING A BVT PROCEDURE, THE ANASTOCLIP GC CLIP APPLIER JAMMED AND STUCK ONTO THE PATIENT'S ARM AND COULD NOT GET IT TO RELEASE THE CLIP OR OFF THE PATIENT'S ARM. THEY HAD TO CUT OUT PART OF THE VEIN. DEVICE WAS CHECKED PRIOR TO USE. A REPLACEMENT DEVICE WAS USED TO COMPLETE THE OPERATION. NO OTHER ISSUES OR INJURY WERE REPORTED TO THE PATIENT.
Additional Manufacturer Narrative
THE DEVICE WAS RECEIVED FOR INVESTIGATION. DURING TESTING THE DEVICE, THE CLIPS WERE ABLE TO CLOSE COMPLETELY AND DEPLOYED PROPERLY FROM THE DEVICE. NO ISSUE OCCURRED DURING TESTING. THE CAUSE OF THE ISSUE DURING THE PROCEDURE COULD NOT BE DETERMINED. IT IS POSSIBLE THAT AN INTERMITTENT ISSUE CAUSED THE CLIP TO BECOME STUCK DURING THE PROCEDURE. DUE TO SIMILAR REPORTED ISSUES, CAPA 2025-004 HAS BEEN INITIATED TO FURTHER INVESTIGATE AND ADDRESS THE ISSUE. NO FURTHER ACTION IS REQUIRED. THE LOT HISTORY RECORD FOR THE DEVICES WAS REVIEWED, NO ISSUES WERE FOUND DURING MANUFACTURING OR PACKAGING THAT WOULD CAUSE OR CONTRIBUTE TO THE REPORTED EVENT. WE HAVE NOT RECEIVED ANY OTHER COMPLAINTS OF A SIMILAR NATURE FOR DEVICES FROM THIS LOT.