MALLINCKRODT
Received Aug 5, 2025 · Event occurred Jul 22, 2025
Report 8020889-2025-00348 · MDR key 22720069
Device
Generic name
Tube, Bronchial (w/wo Connector)
Manufacturer
Mallinckrodt MedicalModel number
126-35Catalog number
126-35
Lot number
202103048X
Product problems
- Use of Device Problem
- Use of Device Problem
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
ADDITIONAL INFORMATION: D9, G3, G3 AND H6. H3 EVALUATION SUMMARY: MEDTRONIC CONDUCTED AN INVESTIGATION BASED UPON ALL INFORMATION RECEIVED. THE DEVICE WAS AVAILABLE FOR EVALUATION. THE RETURNED UNIT WAS INFLATED AND THE TRACHEAL CUFF WOULD NOT INFLATE. IT WAS REPORTED THAT AIR LEAKAGE OCCURRED FROM THE CUFF OF A DOUBLE-LUMEN BRONCHIAL TUBE AT A HOSPITAL. THE REPORTED ISSUE WAS CONFIRMED. THE PRODUCT ANALYSIS NOTED EVIDENCE THAT THE DEVICE WAS NOT USED AS INTENDED. THE MANUFACTURING RECORDS FOR EACH DEVICE ARE THOROUGHLY REVIEWED PRIOR TO RELEASE TO ENSURE THAT IT MEETS ALL MEDTRONIC QUALITY SPECIFICATIONS. THE INSTRUCTIONS INCLUDED WITH THIS DEVICE PROVIDE THE FOLLOWING GUIDANCE: VARIOUS BONY ANATOMICAL STRUCTURES (E.G., TEETH) WITHIN THE INTUBATION ROUTE OR ANY INTUBATION TOOLS WITH SHARP SURFACES PRESENT A THREAT TO MAINTAINING CUFF INTEGRITY. CARE MUST BE TAKEN TO AVOID DAMAGING THE THIN-WALLED CUFFS DURING INTUBATION WHICH WOULD CREATE THE NEED TO SUBJECT THE PATIENT TO THE TRAUMA OF EXTUBATING AND RE-INTUBATION. IF EITHER CUFF IS DAMAGED, THE TUBE SHOULD NOT BE USED MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
Description of Event or Problem
IT WAS REPORTED THAT AIR LEAKAGE OCCURRED FROM THE CUFF OF A DOUBLE-LUMEN BRONCHIAL TUBE AT A HOSPITAL. THERE WAS NO PATIENT INVOLVED.
Additional Manufacturer Narrative
MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.