Received Aug 4, 2025 · Event occurred Jun 20, 2025

Report 2024168-2025-07972 · MDR key 22702007

Device

Generic name

Coronary Stent Delivery System

Model number

1012817-19C

Catalog number

1012817-19C

Lot number

4091641

Product problems

  • Failure to Advance
  • Failure to Advance

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A PERFORATION IN THE LEFT CIRCUMFLEX CORONARY ARTERY WITH MODERATE CALCIFICATION AND MODERATE TORTUOSITY. THE 2.8X19 MM GRAFTMASTER COVERED STENT FAILED TO CROSS DUE TO THE ANATOMY. AN UNSPECIFIED STENT WAS USED TO TREAT THE AREA. THERE WERE NO ADVERSE PATIENT SEQUELA. NO ADDITIONAL INFORMATION WAS PROVIDED.

Additional Manufacturer Narrative

MANUFACTURER'S INVESTIGATION IS STILL PENDING AT THIS TIME. RESULTS AND CONCLUSIONS WILL BE PROVIDED IN THE FINAL REPORT.

Additional Manufacturer Narrative

VISUAL INSPECTION AND DIMENSIONAL ANALYSIS WERE PERFORMED ON THE RETURNED DEVICE. THE REPORTED FAILURE TO ADVANCE COULD NOT BE REPLICATED IN A TESTING ENVIRONMENT AS IT WAS RELATED TO OPERATIONAL CONTEXT OF THE PROCEDURE. PRODUCTION RECORD AND CORRECTIVE AND PREVENTATIVE ACTIONS (CAPA) REVIEWS WERE PERFORMED AND REVEALED NO INDICATION OF A PRODUCT QUALITY ISSUE. ADDITIONALLY, A QUERY OF THE COMPLAINT HANDLING DATABASE FOR THE REPORTED LOT REVEALED THERE IS NO INDICATION OF A LOT OF SPECIFIC ISSUE. BASED ON THE REPORTED INFORMATION AND THE OBSERVATIONS FROM THE RETURNED ANALYSIS, THE INVESTIGATION DETERMINED THAT THE REPORTED ISSUE APPEARS TO BE RELATED TO CIRCUMSTANCES OF THE PROCEDURE. BASED ON THE RESULTS OF THE COMPLAINT INVESTIGATION THERE IS NO INDICATION OF A PRODUCT QUALITY ISSUE WITH RESPECT TO THE DESIGN, MANUFACTURE, OR LABELING OF THE DEVICE.