inforMED
MalfunctionCBL

MEDISORB

Received Aug 1, 2025 · Event occurred Jun 26, 2025

Report 3010838917-2025-00008 · MDR key 22692402

Device

Generic name

Medisorb¿ Ex, Disposable

Manufacturer

Airlife Finland Oy

Model number

2079796-001

Catalog number

2079796-001

Lot number

L01A-02213

Product problems

  • No Flow
  • No Pressure
  • No Flow
  • No Pressure

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE PRODUCT INVOLVED IN THE REPORT HAS NOT BEEN RETURNED. A REVIEW OF THE DEVICE HISTORY RECORD IS IN-PROGRESS. ALL INFORMATION REASONABLY KNOWN AS OF 01 AUG 2025 HAS BEEN INCLUDED IN THE HEALTH AUTHORITY REPORT. SHOULD ADDITIONAL INFORMATION BE OBTAINED A FOLLOW-UP HEALTH AUTHORITY REPORT WILL BE PROVIDED. THE INFORMATION PROVIDED BY AIRLIFE REPRESENTS ALL OF THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT/REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO AIRLIFE. AIRLIFE HAS NO INDEPENDENT KNOWLEDGE OF THE EVENT REPORTED BUT IS RELAYING THE INFORMATION THAT WAS PROVIDED BY THE USER FACILITY WHERE THE INCIDENT OCCURRED. THIS PRODUCT INCIDENT IS DOCUMENTED IN THE AIRLIFE COMPLAINT DATABASE AND IS IDENTIFIED AS COMPLAINT - (B)(4). THIS INFORMATION IS SUBMITTED PURSUANT TO 21CFR803, IN COMPLIANCE WITH THE MEDICAL DEVICE REPORTING REQUIREMENT AND SHOULD NOT BE CONSIDERED TO BE AN ADMISSION THAT A AIRLIFE PRODUCT IS DEFECTIVE OR CAUSES SERIOUS INJURY. ANAESTHETIC MACHINE FAULT - CIRCUIT WILL NOT PRESSURIZE. UNABLE TO RESOLVE THEREFORE ENGINEER CONTACTED WHO INVESTIGATED FAULT. RESULT: FAULTY MEDISORB CANNISTER - NO VISIBLE SIGNS OF ANY DAMAGE. ONCE CANNISTER RENEWED MACHINE IN WORKING ORDER. NUMBER OF PATIENTS INVOLVED: 0 . THE COMPLAINT OF "ANAESTHETIC MACHINE FAULT - CIRCUIT WILL NOT PRESSURIZE. " REGARDING PART 2079796-001 WAS NOT CONFIRMED. THE ROOT CAUSE WAS NOT DETERMINED. A RISK ASSESSMENT WAS PERFORMED, AND THE ULTIMATE RISK WAS DETERMINED TO BE LOW WHICH DOES NOT REQUIRE ESCALATION TO THE CAPA REVIEW BOARD. THERE HAVE BEEN 0 OTHER COMPLAINTS REGARDING THE SAME PART AND A SIMILAR ISSUE WITHIN THE 24 MONTHS PRECEDING THIS REPORTED EVENT. A RESOLUTION LETTER WAS SENT TO THE CUSTOMER. COMPLAINTS WILL CONTINUE TO BE MONITORED FOR POSSIBLE TRENDS.

Description of Event or Problem

IT WAS REPORTED THAT THERE WAS AN ANESTHETIC MACHINE FAULT. THE CIRCUIT WOULD NOT PRESSURIZE. IT IS UNKNOWN IF THERE WAS PATIENT INJURY, HARM OR DELAY IN THERAPY, ALTHOUGH, IF THIS MALFUNCTION WERE TO RECUR DURING PATIENT USE THERE IS A POSSIBILITY OF DELAY IN TREATMENT, SERIOUS INJURY OR HARM TO THE PATIENT.

Additional Manufacturer Narrative

THE PRODUCT INVOLVED IN THE REPORT HAS NOT BEEN RETURNED. A REVIEW OF THE DEVICE HISTORY RECORD IS IN-PROGRESS. ALL INFORMATION REASONABLY KNOWN AS OF 01 AUG 2025 HAS BEEN INCLUDED IN THE HEALTH AUTHORITY REPORT. SHOULD ADDITIONAL INFORMATION BE OBTAINED A FOLLOW-UP HEALTH AUTHORITY REPORT WILL BE PROVIDED. THE INFORMATION PROVIDED BY AIRLIFE REPRESENTS ALL OF THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT/REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO AIRLIFE. AIRLIFE HAS NO INDEPENDENT KNOWLEDGE OF THE EVENT REPORTED BUT IS RELAYING THE INFORMATION THAT WAS PROVIDED BY THE USER FACILITY WHERE THE INCIDENT OCCURRED. THIS PRODUCT INCIDENT IS DOCUMENTED IN THE AIRLIFE COMPLAINT DATABASE AND IS IDENTIFIED AS COMPLAINT: (B)(4). THIS INFORMATION IS SUBMITTED PURSUANT TO 21CFR803, IN COMPLIANCE WITH THE MEDICAL DEVICE REPORTING REQUIREMENT AND SHOULD NOT BE CONSIDERED TO BE AN ADMISSION THAT A AIRLIFE PRODUCT IS DEFECTIVE OR CAUSES SERIOUS INJURY.

Description of Event or Problem

IT WAS REPORTED THAT THERE WAS AN ANESTHETIC MACHINE FAULT. THE CIRCUIT WOULD NOT PRESSURIZE. IT IS UNKNOWN IF THERE WAS PATIENT INJURY, HARM OR DELAY IN THERAPY, ALTHOUGH, IF THIS MALFUNCTION WERE TO RECUR DURING PATIENT USE THERE IS A POSSIBILITY OF DELAY IN TREATMENT, SERIOUS INJURY OR HARM TO THE PATIENT.