inforMED
MalfunctionCEK

TOTAL PROTEIN GEN.2

Received Jul 31, 2025 · Event occurred Jul 3, 2025

Report 1823260-2025-02363 · MDR key 22676035

Device

Generic name

Biuret (colorimetric), Total Protein

Manufacturer

Roche Diagnostics

Catalog number

03183734190

Lot number

850556

Product problems

  • Incorrect Measurement
  • Incorrect Measurement

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

A FIELD SERVICE ENGINEER (FSE) DETERMINED THAT THE REAGENT PACK NEEDED TO BE RECALIBRATED. THE FSE CHECKED THE PROBE ALIGNMENTS, THE GEAR PUMP PRESSURE WAS ADJUSTED, THERE WERE NO LEAKS FOUND, AND THE RINSE HEAD MECHANISM NOZZLES WERE CHECKED AND WERE OKAY. QC AND PRECISION WERE COMPLETED AND PASSED. THE INVESTIGATION DETERMINED THAT THE SERVICE ACTION RESOLVED THE ISSUE.

Additional Manufacturer Narrative

THE COBAS 6000 C (501) MODULE SERIAL NUMBER IS (B)(6). THE INVESTIGATION IS ONGOING.

Description of Event or Problem

THERE WAS AN ALLEGATION OF 12 QUESTIONABLE TOTAL PROTEIN GEN.2 RESULTS FROM THE COBAS 6000 C (501) MODULE. ONLY THE DATA FOR 1 OF THE PATIENTS WAS PROVIDED, AND IS A REPORTABLE MALFUNCTION. THE PATIENT'S INITIAL RESULT WAS 0.60 G/DL, AND THE REPEAT RESULT FROM ANOTHER INSTRUMENT WAS 6.01 G/DL. THE REPEAT RESULTS WERE DEEMED TO BE CORRECT. THE INITIAL RESULTS WERE REPEATED BECAUSE THE DOCTOR OR TECHNICIAN QUESTIONED THE CRITICAL RESULTS.