inforMED
MalfunctionBTQ

RUSCH NASOPHARYNGEAL AIRWAY, LATEX FREE

Received Jul 31, 2025 · Event occurred Mar 31, 2025

Report 8040412-2025-00192 · MDR key 22673799

Device

Generic name

Airway, Nasal

Catalog number

185420-000260

Lot number

KMA21C0058

Product problems

  • Packaging Problem
  • Packaging Problem

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

QN# (B)(4). THE REPORTED COMPLAINT 'TUBES STICKING TO THE PACKAGING' WAS CONFIRMED THROUGH THE INVESTIGATION OF THE RETURNED SAMPLE. THE CUSTOMER RETURNED A BOX CONTAINING FIVE UNOPENED PACKAGES OF SAMPLES FOR INVESTIGATION. TWO OF THE SAMPLES WERE OBSERVED TO HAVE CRYSTALLIZATION. A DEVICE HISTORY RECORD REVIEW WAS NOT REQUIRED FOR THIS INVESTIGATION. BASED ON THE INVESTIGATION FINDINGS, THE COMPLAINT IS CONFIRMED AND VERIFIED TO BE RELATED TO THE MANUFACTURING PROCESS. FURTHER INVESTIGATION HAS BEEN INITIATED UNDER TELEFLEX'S QUALITY SYSTEM BY THE MANUFACTURING SITE TO EVALUATE THIS ISSUE.

Description of Event or Problem

IT WAS REPORTED THAT: "TUBES STICK TO THE PACKAGING".

Additional Manufacturer Narrative

QN#(B)(4).

Description of Event or Problem

IT WAS REPORTED THAT: "TUBES STICK TO THE PACKAGING".