inforMED
MalfunctionEJB

MINIATURE PIN & LIGATURE CUTTER

Received Aug 17, 2011

Report 2086211-2011-00005 · MDR key 2264973

Device

Manufacturer

Gac International

Catalog number

ODG1002

Lot number

03-43

Product problems

  • Material Separation

Patient

Not reported

  • No Consequences Or Impact To Patient

Narrative

Description of Event or Problem

IN THIS EVENT IT WAS REPORTED THAT THE TIP OF A MINIATURE PIN AND LIGATURE CUTTER SEPARATED OUTSIDE OF A PT'S MOUTH; THERE IS NO INDICATION THAT INJURY RESULTED.

Additional Manufacturer Narrative

THE DEVICE WAS RETURNED FOR EVAL AND WAS FOUND TO HAVE EXCEEDED THE EXPECTED LIFE OF THE DEVICE. WHILE NO SERIOUS INJURY RESULTED IN THIS EVENT, THERE HAS BEEN A PREVIOUS REPORT REC'D IN PAST TWO YRS INVOLVING A SIMILAR DEVICE WHERE THIS MALFUNCTION RESULTED IN A SERIOUS INJURY. THEREFORE, THIS EVENT MEETS THE CRITERIA FOR REPORTABILITY PER 21 CFR PART 803.