TPW 24IN 2-0 D/A SH,SKS-3 BWY
Received Jul 29, 2025 · Event occurred May 1, 2025
Report 2210968-2025-08749 · MDR key 22646885
Device
Product problems
- Break
- Break
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
PRODUCT COMPLAINT # (B)(4). DATE SENT TO THE FDA: 7/29/2025. H6 COMPONENT CODE: G07002 - PENDING EVALUATION OF RETURNED DEVICE. THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY ETHICON, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, ETHICON, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE LOT, AND NO NON-CONFORMANCES WERE IDENTIFIED. A DEVICE HAS BEEN RECEIVED; HOWEVER IT HAS NOT YET BEEN EVALUATED. ANY FURTHER INFORMATION DERIVED FROM THE EVALUATION WILL BE SUBMITTED IN A SUPPLEMENTAL 3500A FORM.
Description of Event or Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN (B)(6) 2025 AND SUTURE WAS USED. IT WAS REPORTED THAT THE NEEDLE HAD BEEN FOUND BROKEN IN THE NEWLY DISPENSED SUTURE WHEN PREPARING FOR SURGERY. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Additional Manufacturer Narrative
PRODUCT COMPLAINT # (B)(4). THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY ETHICON INC, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, ETHICON INC OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE. ADDITIONAL INFORMATION: H6, D9, H3 H3 INVESTIGATION SUMMARY: A FAILED SUTURE NEEDLE, LABELED AS (B)(4), WAS SUBMITTED TO HERRERA, STAFFORD AND ASSOCIATES, LLC (HSA) FOR FRACTOGRAPHIC EVALUATION ON JUNE 20, 2025. THE COMPONENT WAS IDENTIFIED AS PRODUCT CODE TPW30. IT WAS REQUESTED THAT HSA ASSESS THE FRACTURE MODE OF THE FAILURE. ACCORDING TO THE INFORMATION, IT WAS REPORTED BREAKAGE NEEDLE PRE-OP. THE PRODUCT RECEIVED FOR ANALYSIS WAS TPW30. VISUAL INSPECTION AND METALLURGICAL ANALYSIS WERE PERFORMED ON THE RETURNED PRODUCT. A FRACTURE WAS OBSERVED IN THE BODY OF THE NEEDLE. A SCANNING ELECTRON MICROSCOPE (SEM) WAS USED TO EXAMINE THE FRACTURE SURFACES AND SURROUNDING AREA OF THE NEEDLE. THE FRACTURE SURFACES WERE EXAMINED ON SIDE A IN MULTIPLE LOCATIONS TO DETERMINE THE FRACTURE MODE. THE EVALUATION OF (B)(4) REVEALED THE FRACTURE WAS PREDOMINANTLY COMPOSED OF MICROVOID COALESCENCE, WHICH IS EVIDENCE OF A DUCTILE FRACTURE MODE. NECKING OBSERVED ADJACENT TO THE FRACTURE PROVIDES ADDITIONAL EVIDENCE THAT THE FAILURE WAS INDUCED BY MECHANICAL DEFORMATION LEADING TO DUCTILE OVERLOAD. THIS WAS A DUCTILE FRACTURE. ADDITIONAL COMPLAINT INFORMATION MONITORING FOR POTENTIAL SAFETY SIGNALS IS CONDUCTED THROUGH COMPLAINT TRENDING AS PART OF POST-MARKET SURVEILLANCE.