PENTAX
Received Jul 25, 2025 · Event occurred Jul 15, 2025
Report 9610877-2025-52653 · MDR key 22606250
Device
Generic name
Fiberopticintubationscope 1.4c 3.4tp Rop
Manufacturer
Hoya Corporation Pentax Tokyo OfficeModel number
FI-10RBSProduct problems
- Loose or Intermittent Connection
- Loose or Intermittent Connection
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
THE TIME OF EVENT IS NOT DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT HARM. SUCTION NIPPLE LOOSENED.
Additional Manufacturer Narrative
G4: THIS DEVICE IS NOT DISTRIBUTED IN US SO THAT 510K# IS BLANK. D4 UDI: "NOT DISTRIBUTED IN USA", BECAUSE PRODUCTS ARE NOT DISTRIBUTED IN USA PRODUCTS, BUT SIMILAR DEVICE PRODUCT ARE LISTED IN USA. THE PRODUCT WAS RETURNED TO PENTAX MEDICAL FOR REPAIR. OUR TECHNICIAN CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE SUCTION NIPPLE LOOSE. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE EXCESSIVE FORCE APPLIED ON THE SUCTION NIPPLE. IN ADDITION, OUR TECHNICIAN CONFIRMED THAT THE CFB (COHERENT FIBER BUNDLE) BROKEN, THE LCB (LIGHT CARRYING BUNDLE) BROKEN, AND THE INSERTION FLEXIBLE TUBE CUT; HOWEVER, THESE DEFECTS ARE NOT THE MAIN CAUSE, AND/OR IRRELEVANT TO THE ALLEGED COMPLAINT. THIS DEFECT OCCURRED NOT DURING PROCEDURE. BASED ON THE TECHNICAL REPORT ""HR-RPT-0588 (CHANNEL)"" AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.