inforMED
MalfunctionCEC

ELECSYS FT4

Received Jul 24, 2025

Report 1823260-2025-02263 · MDR key 22587302

Device

Generic name

Free Thyroxine

Manufacturer

Roche Diagnostics

Model number

FT4 IV

Catalog number

ASKU

Lot number

ASKU

Product problems

  • False Positive Result
  • False Positive Result

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

1. SUMMARY OF INVESTIGATION RESULTS: THE INVESTIGATION IS ONGOING. 2. SUMMARY OF FOLLOW UP/CORRECTIVE ACTIONS TAKEN: THE INVESTIGATION IS ONGOING. 3. DEVICE RETURNED TO MANUFACTURER? NO. 4. DEVICE LABELED "FOR SINGLE USE?" NO. 5. DEVICE REPROCESSED AND REUSED? NO.

Description of Event or Problem

1. THERE WAS AN ALLEGATION OF QUESTIONABLE ELECSYS FT4 RESULTS FROM THE COBAS E 801 ANALYTICAL UNIT.

Additional Manufacturer Narrative

1. SUMMARY OF INVESTIGATION RESULTS: THE INVESTIGATION DETERMINED THE ISSUE WAS DUE TO PRE-ANALYTICAL HANDLING ISSUES AT THE CUSTOMER SITE. CORRECT PRE-ANALYTIC SAMPLE HANDLING IS WITHIN THE CUSTOMER'S RESPONSIBILITY. THERE IS NO EVIDENCE TO SUGGEST A MALFUNCTION OF THE DEVICE. 2. SUMMARY OF FOLLOW UP/CORRECTIVE ACTIONS TAKEN: NONE.