inforMED
MalfunctionPEO

EPLEX BLOOD CULTURE IDENTIFICATION FUNGAL PATHOGEN (BCID-FP) PANEL

Received Jul 24, 2025 · Event occurred Jun 25, 2025

Report 3008632402-2025-02245 · MDR key 22586086

Device

Generic name

Fungal Organisms, Nucleic Acid-based Assay

Catalog number

09556516001

Lot number

10223348

Product problems

  • False Negative Result
  • False Positive Result
  • False Negative Result
  • False Positive Result

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE COBAS EPLEX CORE UNIT SERIAL NUMBER WAS (B)(6). THE INVESTIGATION IS ONGOING.

Description of Event or Problem

THERE WAS AN ALLEGATION OF QUESTIONABLE RESULTS FOR THE EPLEX BLOOD CULTURE IDENTIFICATION FUNGAL PATHOGEN (BCID-FP) PANEL USING THE COBAS EPLEX CORE UNIT. SAMPLE 1 INITIAL RESULT WAS C. ALBICANS AND C. FAMATA. ON (B)(6) 2025, THE REPEAT RESULT WAS C. ALBICANS AND C. FAMATA. THE CULTURE RESULTS WERE C. ALBICANS AND C. DUBLINIENSIS SAMPLE 2 INITIAL RESULT WAS C. DUBLINIENSIS AND C. FAMATA. ON (B)(6) 2025, THE REPEAT RESULT WAS C. FAMATA. THE CULTURE RESULT WAS C. DUBLINIENSIS.

Additional Manufacturer Narrative

THE INVESTIGATION DID NOT IDENTIFY A SPECIFIC ROOT CAUSE.