inforMED
MalfunctionCEC

ALINITY I FREE T4 REAGENT KIT

Received Jul 22, 2025 · Event occurred Jul 6, 2025

Report 3005094123-2025-00367 · MDR key 22568339

Device

Generic name

Radioimmunoassay, Free Thyroxine

Catalog number

07P70-20

Lot number

68517UD00

Product problems

  • High Test Results
  • High Test Results

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

THE CUSTOMER OBSERVED FALSELY ELEVATED ALINTY I FREE T4 RESULTS FOR ONE PATIENT. THE FOLLOWING RESULTS WERE PROVIDED: SID (B)6), FIRST VALUE 19.80, REPEAT RUN 14.71 PMOL/L. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

Additional Manufacturer Narrative

THE COMPLAINT INVESTIGATION INCLUDED A REVIEW OF DATA AND INFORMATION PROVIDED BY THE CUSTOMER, SEARCH FOR SIMILAR COMPLAINTS, TICKET TRENDING REVIEW, DEVICE HISTORY RECORD REVIEW, IN HOUSE TESTING, AND LABELING REVIEW. DATA AND INFORMATION PROVIDED BY THE CUSTOMER WERE REVIEWED AND SUPPORT THE COMPLAINT ISSUE. REVIEW OF TRACKING AND TRENDING DATA FOR THE ALINITY I FREE T4 ASSAY DID IDENTIFY AN INCREASE IN COMPLAINTS. ADDITIONALLY, AN INCREASE IN COMPLAINT ACTIVITY WAS IDENTIFIED FOR REAGENT LOT 68517UD00. HOWEVER, IN HOUSE TESTING WAS PERFORMED UTILIZING RETAINED KITS OF LOT 68517UD00. ALL TESTING PASSED INDICATING THE REAGENT LOT IS PERFORMING AS EXPECTED. A DEVICE HISTORY RECORD REVIEW DID NOT IDENTIFY ANY NONCONFORMANCES, POTENTIAL NONCONFORMANCES OR DEVIATIONS ASSOCIATED WITH THE COMPLAINT ISSUE. LABELING WAS REVIEWED WHICH ADEQUATELY ADDRESSES THE CURRENT ISSUE. BASED ON THE INVESTIGATION, NO SYSTEMIC ISSUE OR DEFICIENCY OF THE ALINITY I FREE T4 REAGENT LOT 68517UD00 WAS IDENTIFIED.

Description of Event or Problem

THE CUSTOMER OBSERVED FALSELY ELEVATED ALINTY I FREE T4 RESULTS FOR ONE PATIENT. THE FOLLOWING RESULTS WERE PROVIDED: SID: (B)(6) FIRST VALUE 19.80, REPEAT RUN 14.71 PMOL/L. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

Additional Manufacturer Narrative

AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE.