ALINITY I CA 19-9XR REAGENT KIT
Received Jul 22, 2025 · Event occurred Jul 5, 2025
Report 3002809144-2025-00241 · MDR key 22564361
Device
Generic name
System, Test, Carbohydrate Ag (ca19-9), For Monit And Mgmt Of Pancreatic Cancer
Manufacturer
Abbott GmbhCatalog number
08P32-20
Lot number
71456FP00
Product problems
- Incorrect, Inadequate or Imprecise Result or Readings
- Incorrect, Inadequate or Imprecise Result or Readings
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
CORRECTION: SECTION H11: TYPOGRAPHICAL ERROR IN THE PRODUCT EVALUATION, SEE CORRECTED PRODUCT EVALUATION: THE COMPLAINT INVESTIGATION FOR IMPRECISE ALINITY I CA 19-9XR RESULTS INCLUDED A REVIEW OF DATA AND INFORMATION PROVIDED BY THE CUSTOMER, SEARCH FOR SIMILAR COMPLAINTS, TICKET TRENDING REVIEW, LABELING REVIEW, DEVICE HISTORY RECORD REVIEW, AND FIELD DATA REVIEW. RETURN TESTING WAS NOT COMPLETED AS RETURNS WERE NOT AVAILABLE. DATA AND INFORMATION PROVIDED BY THE CUSTOMER WERE REVIEWED AND SUPPORTED THE COMPLAINT ISSUE. REVIEW OF TRACKING AND TRENDING FOR THE ALINITY I CA 19-9XR ASSAY DID NOT IDENTIFY AN INCREASE IN COMPLAINT ACTIVITY RELATED TO THE COMPLAINT ISSUE. ADDITIONALLY, AN INCREASE IN COMPLAINT ACTIVITY WAS NOT IDENTIFIED FOR REAGENT LOT 71456FP00. THE DEVICE HISTORY RECORD REVIEW DID NOT IDENTIFY ANY NON-CONFORMANCES, DEVIATIONS, OR POTENTIAL NON-CONFORMANCES. LABELING WAS REVIEWED AND SUFFICIENTLY ADDRESSES THE CUSTOMER'S ISSUE.THE OVERALL PERFORMANCE OF ALINITY I CA 19-9XR WAS REVIEWED USING FIELD DATA FROM CUSTOMERS WORLDWIDE. THE PATIENT¿S DATA WAS ANALYZED AND COMPARED TO AN ESTABLISHED CONTROL LIMIT. THIS EVALUATION INDICATED THAT THE PATIENT MEDIAN RESULT FOR LOT 71456FP00 IS WITHIN THE ESTABLISHED CONTROL LIMITS, THEREFORE, NO UNUSUAL REAGENT LOT PERFORMANCE WAS IDENTIFIED FOR 71456FP00. BASED ON THE INVESTIGATION, NO SYSTEMIC ISSUE OR DEFICIENCY WITH THE ALINITY I CA 19-9XR ASSAY FOR LOT 71456FP00 WAS IDENTIFIED. ALL AVAILABLE PATIENT INFORMATION WAS INCLUDED. ADDITIONAL PATIENT DETAILS ARE NOT AVAILABLE.
Description of Event or Problem
THE CUSTOMER OBSERVED IMPRECISE ALINITY I CA 19-9XR RESULTS FOR ONE PATIENT. THE CUSTOMER OBSERVED A DIFFERENCE IN THE INITIAL TEST AND DILUTED RETEST RESULTS. THE CUSTOMER REPORTED BOTH SAMPLE IDS ARE FOR THE SAME PATIENT. THE FOLLOWING RESULTS WERE PROVIDED: (B)(6) 2025 SID (B)(6) INITIAL RESULT = 1375.368 U/ML, REPEAT RESULT WITH AUTOMATIC DILUTION = 779.982 U/ML, SID (B)(6) 1188.324 U/ML, RESULT WITH AUTOMATIC DILUTION = 603.267 U/ML . NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Description of Event or Problem
THE CUSTOMER OBSERVED IMPRECISE ALINITY I CA 19-9XR RESULTS FOR ONE PATIENT. THE CUSTOMER OBSERVED A DIFFERENCE IN THE INITIAL TEST AND DILUTED RETEST RESULTS. THE CUSTOMER REPORTED BOTH SAMPLE IDS ARE FOR THE SAME PATIENT. THE FOLLOWING RESULTS WERE PROVIDED: (B)(6) 2025 SID (B)(6) INITIAL RESULT = 1375.368 U/ML, REPEAT RESULT WITH AUTOMATIC DILUTION = 779.982. U/ML, SID (B)(6) 1188.324 U/ML, RESULT WITH AUTOMATIC DILUTION = 603.267 U/ML. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Additional Manufacturer Narrative
THE COMPLAINT INVESTIGATION FOR IMPRECISE ALINITY I CA 19-9XR RESULTS INCLUDED A REVIEW OF DATA AND INFORMATION PROVIDED BY THE CUSTOMER, SEARCH FOR SIMILAR COMPLAINTS, TICKET TRENDING REVIEW, LABELING REVIEW, DEVICE HISTORY RECORD REVIEW, AND FIELD DATA REVIEW. RETURN TESTING WAS NOT COMPLETED AS RETURNS WERE NOT AVAILABLE. DATA AND INFORMATION PROVIDED BY THE CUSTOMER WERE REVIEWED AND SUPPORTED THE COMPLAINT ISSUE. REVIEW OF TRACKING AND TRENDING FOR THE ALINITY I CA 19-9XR ASSAY DID NOT IDENTIFY AN INCREASE IN COMPLAINT ACTIVITY RELATED TO THE COMPLAINT ISSUE. ADDITIONALLY, AN INCREASE IN COMPLAINT ACTIVITY WAS NOT IDENTIFIED FOR REAGENT LOT 71456FP00. THE DEVICE HISTORY RECORD REVIEW DID NOT IDENTIFY ANY NON-CONFORMANCES, DEVIATIONS, OR POTENTIAL NON-CONFORMANCES. LABELING WAS REVIEWED AND SUFFICIENTLY ADDRESSES THE CUSTOMER'S ISSUE. TESTING WAS PERFORMED TO EVALUATE THE PERFORMANCE OF THE REAGENT LOT 71456FP00. THE OVERALL PERFORMANCE OF ALINITY I CA 19-9XR WAS REVIEWED USING FIELD DATA FROM CUSTOMERS WORLDWIDE. THE PATIENT¿S DATA WAS ANALYZED AND COMPARED TO AN ESTABLISHED CONTROL LIMIT. THIS EVALUATION INDICATED THAT THE PATIENT MEDIAN RESULT FOR LOT 71456FP00 IS WITHIN THE ESTABLISHED CONTROL LIMITS, THEREFORE, NO UNUSUAL REAGENT LOT PERFORMANCE WAS IDENTIFIED FOR 71456FP00. BASED ON THE INVESTIGATION, NO SYSTEMIC ISSUE OR DEFICIENCY WITH THE ALINITY I CA 19-9XR ASSAY FOR LOT 71456FP00 WAS IDENTIFIED. THIS REPORT IS BEING FILED ON AN INTERNATIONAL PRODUCT, LIST NUMBER 08P32-20 THAT HAS A SIMILAR PRODUCT DISTRIBUTED IN THE US, LIST NUMBER 8P32-21 WITH 510K/PMA/BLA NUMBER K052000 ALL AVAILABLE PATIENT INFORMATION WAS INCLUDED. ADDITIONAL PATIENT DETAILS ARE NOT AVAILABLE.